Z-3009-2011

Siemens Axiom Artis U mobile x-ray system

Siemens Medical Solutions USA, Inc. / initiated 2011-06-29 / Terminated / root cause: labeling and UDI

Siemens Medical Solutions USA, Inc began a recall of Siemens Axiom Artis U mobile x-ray system Product Usage: Mobile x-ray system on . The reason FDA records is "When the heat unit of the angiography system reaches or exceeds 65%, the cool down time display will appear with an acoustic signal. If users ignore the warning and continue the radiation, the tube may be autoprotected...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-3009-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Siemens Axiom Artis U mobile x-ray system Product Usage: Mobile x-ray system
Product code
IZL System, X-Ray, Mobile
Reason for recall
When the heat unit of the angiography system reaches or exceeds 65%, the cool down time display will appear with an acoustic signal. If users ignore the warning and continue the radiation, the tube may be autoprotected when it reaches its temperature limitation causing the X-Ray to lock until the tube cools down. Following a tube auto-protection, an error message "Door is Open" or "Err 421/240:
Root cause tag
labeling and UDI
FDA root cause
Labeling False and Misleading
Action
Siemens sent a "Customer Safety Advisory Notice" letters to all affected users dated June 13, 2011. The letter identified the product, the problem, and the actions to be taken. Siemens included in the notice an addendum to the instruction for use to each system and request customers to take special attention. Customers were instructed to place the Safety Notice in the System's instruction for use. All appropriate personnel should be notified of the correction notice. If product was distributed further, customers were to forward the safety notice to the new owner and inform the firm of the new owner.
Quantity in commerce
18
Distribution
Nationwide Distribution (USA) including the states of: CA, CT, FL, KS, MN, MO, NJ, NY, NC, PA, and UT.
Product codes and lots
Model number 10094012; serial numbers 20202, 20223, 20231, 20422, 20402, 20435, 21022, 22002, 21005, and 20432.   Model number 7008530; serial numbers 20146, 21112, 20003, 20115, 20135, 20006, 20113, and 20120.