Synthes T1 Matrix Locking Cap of the Synthes Matrix System, Part Number 04.632.000. Intended for posterior...
Synthes USA (HQ), Inc. began a recall of Synthes T1 Matrix Locking Cap of the Synthes Matrix System, Part Number 04.632.000. Intended for posterior pedicle screw fixation, posterior hook fixation, or... on . The reason FDA records is "Labeling Correction -- The locking cap was not able to be loosened, resulting in the breakage of the screwdriver or the rod to be cut and the assembly to be removed (screw, cap, and rod). This poses a potential risk of...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-3207-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Synthes GmbH
- Product
- Synthes T1 Matrix Locking Cap of the Synthes Matrix System, Part Number 04.632.000. Intended for posterior pedicle screw fixation, posterior hook fixation, or anterolateral fixation.
- Product code
- NKB Thoracolumbosacral Pedicle Screw System
- Reason for recall
- Labeling Correction -- The locking cap was not able to be loosened, resulting in the breakage of the screwdriver or the rod to be cut and the assembly to be removed (screw, cap, and rod). This poses a potential risk of significant prolongation of OR time.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling Change Control
- Action
- The firm initiated this action on March 11, 2011 by issuing an Urgent: Medical Device Labeling Correction letter to all consignees. The letter identified the affected product and stated the reason for the correction. It also instructed the consignee to review and attach the supplementary technique page to any MATRIX Spine System Technique Guides that they may have had in their facility. Customers are to complete the Verification Section of the letter and return it to the firm via mail, fax, or e-mail. If customers have questions, they can call 610-719-5063 or their Synthes Spine Sales Consultant.
- Distribution
- Nationwide Distribution -- AL, AZ, AR, CA, CO, CT, DE, FL, GA, ID, IL, IN, IA, KS, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WV, and WI.
- Product codes and lots
- All lots