Smith & Nephew, Tendon Stripper, Slotted, 5.0 mm I.D. x 12 working length. Non-Sterile Catalog Number: 013554...
Smith & Nephew, Inc. Endoscopy Division began a recall of Smith & Nephew, Tendon Stripper, Slotted, 5.0 mm I.D. x 12 working length. Non-Sterile Catalog Number: 013554 Orthopedic Manual Surgical Instrument on . The reason FDA records is "Shaft tip (casting) does not meet the SNE 416 specification, which could result in rusting of the Tendon Strippers". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-3050-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Smith & Nephew, Inc.
- Product
- Smith & Nephew, Tendon Stripper, Slotted, 5.0 mm I.D. x 12 working length. Non-Sterile Catalog Number: 013554 Orthopedic Manual Surgical Instrument
- Product code
- LXH Orthopedic Manual Surgical Instrument
- Reason for recall
- Shaft tip (casting) does not meet the SNE 416 specification, which could result in rusting of the Tendon Strippers
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Smith & Nephew sent an Urgent - Product Recall Notification letter dated July 7, 2011. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to complete the form included in the letter and return to [email protected] or fax to 508-261-3620. Each affiliate must then notify each facility affected by the recall. To return product and obtain credit or order replacement product customers should contact the Smith & Nephew Endoscopy Returns at 800-343-5717 (option 3) for a Return Authorization (RA) or send an email to, [email protected]. For questions, contact Hoangthi by phone at 508-337-3731.
- Quantity in commerce
- 570 units
- Distribution
- Worldwide Distribution - USA (nationwide) and the countries of Australia, Austria, Belgium, Canada, Chile, China, Columbia, Denmark, Dubai, Finland, France, Germany, Greece, Hong Kong, India, Italy, Israel, Japan, Korea, Malaysia, Mexico, Netherland, Norway, Singapore, South Africa, Spain, Sweden, Switzerland, Thailand, United Kingdom, and Venezuela
- Product codes and lots
- Lot Numbers: 50342499, 50344799, 50347017, 50349339 50354170, 50357485, 50360691, 50360992 50367247, 50374544, 50375820, 50370949 50371944