Z-2954-2011

IL GEM PCL/PCL Plus - Gem Test PT Cartridge ITC-Nexus Dx A unitized mlcrocoagulation test intended for use in...

Accriva Diagnostics, Inc. / initiated 2011-05-10 / Terminated / root cause: design

International Technidyne Corp. began a recall of IL GEM PCL/PCL Plus - Gem Test PT Cartridge ITC-Nexus Dx A unitized mlcrocoagulation test intended for use in performing one-stage prothrombin time on . The reason FDA records is "Certain lots of Hemochron Jr. PT Cuvettes (J201) and GEM Test PT cartridges demonstrate a higher bias than historically observed when compared to laboratory reference. instruments". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2954-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Accriva Diagnostics, Inc.
Product
IL GEM PCL/PCL Plus - Gem Test PT Cartridge ITC-Nexus Dx A unitized mlcrocoagulation test intended for use in performing one-stage prothrombin time.
Product code
JPA System, Multipurpose For In Vitro Coagulation Studies
Reason for recall
Certain lots of Hemochron Jr. PT Cuvettes (J201) and GEM Test PT cartridges demonstrate a higher bias than historically observed when compared to laboratory reference. instruments.
Root cause tag
design
FDA root cause
Device Design
Action
International Technidyne Corp. (ITC) sent "Urgent Medical Device Recall" letters dated May 10, May 25 and June 09, 2011 to all affected customers. The letter included a description of the problem, asked customers fill out the attached form, and return affected product for replacement. For questions, please contact ITC Technical Support at (800) 631-5945.
Quantity in commerce
265 boxes total
Distribution
Worldwide Distribution: (USA) Nationwide Distribution including the countries of; Austria, Australia, Canada, Switzerland, Germany, Denmark, Estonia, Spain, Finland, France, United Kingdom, Hong Kong, Ireland, Israel, India, Italy, Korea, Liechtenstien, Libya, Netherlands, Norway, Oman,Portugal, Qatar, Saudi Arabia, Sweden, Turkey, Taiwan, and South Africa.
Product codes and lots
ID #: 6260030000.  Lot Numbers:  A1ILJPT001, H0ILJPT005, and K0ILJPT006.