Z-2933-2011
RUSCH, GB: Female Catheter, All Purpose PVC, 10 Fr. 18 cm, Distributed by Teleflex Medical, Research Triangle...
Teleflex Medical / initiated 2011-06-30 / Terminated / root cause: packaging
Teleflex Medical began a recall of RUSCH, GB: Female Catheter, All Purpose PVC, 10 Fr. 18 cm, Distributed by Teleflex Medical, Research Triangle Park, NC Product Usage: Catheter that is inserted... on . The reason FDA records is "Pin holes were detected in the packaging therefore sterility of the device can not be guaranteed". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-2933-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Teleflex Medical
- Product
- RUSCH, GB: Female Catheter, All Purpose PVC, 10 Fr. 18 cm, Distributed by Teleflex Medical, Research Triangle Park, NC Product Usage: Catheter that is inserted through the urethra and used to pass fluids to or from the urinary tract.
- Product code
- KOD Catheter, Urological
- Reason for recall
- Pin holes were detected in the packaging therefore sterility of the device can not be guaranteed.
- Root cause tag
- packaging
- FDA root cause
- Other
- Action
- Teleflex Medical sent an "URGENT MEDICAL DEVICE RECALL NOTIFICATION" letter dated June 30, 2011 to all affected consignees. The letter described the product, problem and actions to be taken. The letter instructs the customers to discontinue use and quarantine any of the affected products. Customers were asked to complete the attached Recall Acknowledgement Form and return as per the instructions. A separate letter was sent to distributors instructing them to communicate the recall to any of their customers who may have receive the affected products, using the recall letter and form as a template. For questions, contact your local sales representative or call Customer service at 1-866-246-6990
- Quantity in commerce
- See line 1.
- Distribution
- Worldwide Distribution and (USA) Nationwide Distribution.
- Product codes and lots
- Product number: 238300100, Lot numbers: 18754, 19225, 19910, 22276, 22277, 25562, 103029, 105439, 113974, 116419, 118754, 119225, 119910, 122276, 122277, 123172, 125562, 142573, 142574, 310434, 322168, 335515, 340212, 342045, 346326, 350056, 359273, 359990, 378591, 378592, 379600, 380079, 380255, 380289, 384500, 390490, 391728, 405524 and 411563.