INTRABEAM Shielding Material Flat 20 x 20 cm, Sterile/FO 0.05 mm Pb Equivalent (50keV) Carl Zeiss Meditec AG...
Carl Zeiss Meditec, Inc. began a recall of INTRABEAM Shielding Material Flat 20 x 20 cm, Sterile/FO 0.05 mm Pb Equivalent (50keV) Carl Zeiss Meditec AG Indicated for radiation therapy treatments on . The reason FDA records is "Carl Zeiss Meditec received a report that the INTRABEAM 20 x 20 cm sterile shield's packaging looked wet inside. Investigation found the substance to be Triethylene Glycol". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2989-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Carl Zeiss Meditec Inc
- Product
- INTRABEAM Shielding Material Flat 20 x 20 cm, Sterile/FO 0.05 mm Pb Equivalent (50keV) Carl Zeiss Meditec AG Indicated for radiation therapy treatments.
- Product code
- JAD System, Therapeutic, X-Ray
- Reason for recall
- Carl Zeiss Meditec received a report that the INTRABEAM 20 x 20 cm sterile shield's packaging looked wet inside. Investigation found the substance to be Triethylene Glycol,
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Carl Zeiss Meditec, Inc. sent an "Important Field Correction Action" letter dated July 13, 2011 to all affected customers. The letter included description of affected product and problem. Customers are asked to notify the firm if they have experienced any problems with the product and to return any remaining product for replacement upon completion of required testing. Customers are asked to fill out and return an Acknowledgement Form. For additional information contact Carl Zeiss Meditec at (925) 570-4844.
- Quantity in commerce
- 8 boxes of 10 sterile packages.
- Distribution
- Nationwide (USA) Distribution including the states of FL, CA, NY, VA and AL.
- Product codes and lots
- Lot 110851; Catalog number 304534-6000-585.