GammaMed Flexible Applicator probe; 3.2mm diameter, braced PVDF. Varian Medical Systems, Palo Alto, CA. Mfg...
Varian Medical Systems, Inc. Oncology Systems began a recall of GammaMed Flexible Applicator probe; 3.2mm diameter, braced PVDF. Varian Medical Systems, Palo Alto, CA. Mfg by Varian Medical Systems, Haan GmbH. Flexible... on . The reason FDA records is "Certain GammaMed Flexible applicator probes may not have been heat annealed during manufacturing to prevent shrinkage after autoclave". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-3013-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Varian Medical Systems, Inc.
- Product
- GammaMed Flexible Applicator probe; 3.2mm diameter, braced PVDF. Varian Medical Systems, Palo Alto, CA. Mfg by Varian Medical Systems, Haan GmbH. Flexible Applicator Robes are designed for brachytherapy intracavitary treatments in areas such as vagina, vagina stump or rectum.
- Product code
- JAQ System, Applicator, Radionuclide, Remote-Controlled
- Reason for recall
- Certain GammaMed Flexible applicator probes may not have been heat annealed during manufacturing to prevent shrinkage after autoclave.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Varian Medical Systems, Inc. sent out Customer Technical bulletins in August 2008 and Urgent Medical Device Correction/Urgent Field Safety Notices on July 13, 2011 to all affected customers. The letter included description of device and problem. The current notice supplements and modifies previous instructions and ask that they stop using any flexible applicator probes in inventory. They are to be returned to Varian Brachy Therapy for replacement. Letter go on to list possible scenarios if the length deviation is not detected and advised all personnel working in the radiotherapy department should be notified of the recall. For additional information contact (800) 360-7909.
- Quantity in commerce
- 1343
- Distribution
- Worldwide Distribution: Nationwide (USA) and including the countries of France, Germany, India, Trinidad and Tobago, United Kingdom, Canada, Nepal, Ukraine, Czech Republic, Bangladesh, Thailand, Turkey, Brazil, Estonia, Switzerland, Russia, Greece, Philippines, Spain, Belgium, Kuwait, Romania, Morocco, Mexico, Columbia, Austria, Poland, Kazahstan, Vietnam, Cyprus, New Zealand, Dominican Republic, Norway, Uzebestan, Hungary, Lebanon, South Africa, South Korea, Slovenia, Venezuela, Serbia, Tunisia, Pakistan, Macedonia, Ireland, Azerbajian, Ukraine, Morroco, Italy, Singapore, Nambia, Turkey, Sudan, and Taiwan.
- Product codes and lots
- Lots D01 to H14; Part number GM1102560