Supreme Electrophysiology Extension Cable, REF 401984, For use with Hexapolar Catheters, Sterile EO, Rx only...
St. Jude Medical began a recall of Supreme Electrophysiology Extension Cable, REF 401984, For use with Hexapolar Catheters, Sterile EO, Rx only, St Jude Medical, 14901 DeVeau Place, Minnetonka... on . The reason FDA records is "St. Jude Medical is recalling a single lot (3334664) of Supreme Electrophysiology Extension Cable which were packaged with a Quadripolar cable instead of a hexapolar Cable. St. Jude Medical has determined that the...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-2955-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- Supreme Electrophysiology Extension Cable, REF 401984, For use with Hexapolar Catheters, Sterile EO, Rx only, St Jude Medical, 14901 DeVeau Place, Minnetonka, MN 55345. Supreme Electrophysiology Extension Cable used to connect Electrophysiology catheters to external diagnostic or therapeutic equipment. The section indexing connector located a t the distal end of the cable will only connect to SMJ Electrophysiology catheters.
- Product code
- IKD Cable, Electrode
- Reason for recall
- St. Jude Medical is recalling a single lot (3334664) of Supreme Electrophysiology Extension Cable which were packaged with a Quadripolar cable instead of a hexapolar Cable. St. Jude Medical has determined that the incorrectly labeled product poses no safety risk since the extension cable can be safely used with a corresponding Quadripolar catheter but cannot be incorrectly connected to a Hexapolar catheter.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling mix-ups
- Action
- St. Jude Medical sent an "IMPORTANT MEDICAL DEVICE FIELD ADVISORY NOTICE" letter dated June 8, 2011 to all affected customers. The letter describes the product, problem, and the actions to be taken by the customers. The letter instructs customers to locate the affected product and to complete and return an attached SJM AFD Reconciliation Form. The form will be used to replace product and as documentation that the customers received the notice. Fax the completed form to 877-497-031. Contact the Director of Product Marketing at 512-329-1054 for questions regarding this notice.
- Quantity in commerce
- 32
- Distribution
- Nationwide Distribution-USA (nationwide) including the states of FL, NC, NY, and OR.
- Product codes and lots
- Batch # 3334664