Latitude", Tige Humerale, Humeral Stem, Medium 77 mm, Humeral Stem Right, For Cemented use only, Sterile R...

TORNIER Inc. / initiated 2011-06-23 / Terminated / root cause: manufacturing process

Tornier, Inc began a recall of Latitude", Tige Humerale, Humeral Stem, Medium 77 mm, Humeral Stem Right, For Cemented use only, Sterile R, REF 0030402, Tornier In. Edina, MN 55435 USA. The... on . The reason FDA records is "Tornier has discovered that some models of the Tornier latitude Elbow Prosthesis (Small and Medium Humeral stems) were manufactured with a humeral screw that may not function as designed". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-3082-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
TORNIER Inc.
Product
Latitude", Tige Humerale, Humeral Stem, Medium 77 mm, Humeral Stem Right, For Cemented use only, Sterile R, REF 0030402, Tornier In. Edina, MN 55435 USA. The Tornier In. Latitude Elbow Prosthesis is intended for total elbow arthroplasty. Prosthetic replacement with this device may be indicated to relieve severe pain or significant disability following the effects of primary or secondary osteoarthritis and rheumatoid arthritis; correction of functional deformities; revision procedures where other treatments or devices have failed; treatment of fractures that are unmanageable using others techniques. The Tornier In. Latitude Elbow Prosthesis is intended for cemented use only.
Product code
JDB Prosthesis, Elbow, Semi-Constrained, Cemented
Reason for recall
Tornier has discovered that some models of the Tornier latitude Elbow Prosthesis (Small and Medium Humeral stems) were manufactured with a humeral screw that may not function as designed.
Root cause tag
manufacturing process
FDA root cause
Process change control
Action
Tornier, Inc. sent an "Urgent Product Correction" letter dated June 23, 2011. The letter was addressed to the doctors. The letter described the product and the problem. Doctors were insructed to monitor patients that have received the units for evidence of humeral screw loosening. For questions customers were instructed to contact the Customer Service Department at 1-888-494-7950.
Quantity in commerce
3
Distribution
Nationwide Distribution including FL and MO
Product codes and lots
CC0810047