Z-2996-2011
Titanium Fletcher-style Applicator Set - Defined Geometry - CT Compatible with Titanium intrauterine probe...
Varian Medical Systems, Inc. / initiated 2011-07-13 / Terminated / root cause: design
Varian Medical Systems, Inc. Oncology Systems began a recall of Titanium Fletcher-style Applicator Set - Defined Geometry - CT Compatible with Titanium intrauterine probe with stopper length 60 mm, 30 degree angle. Varian... on . The reason FDA records is "Potential of weld failure of the Titanium intrauterine probe at the cervical stopper". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2996-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Varian Medical Systems, Inc.
- Product
- Titanium Fletcher-style Applicator Set - Defined Geometry - CT Compatible with Titanium intrauterine probe with stopper length 60 mm, 30 degree angle. Varian Medical Systems, Palo Alto, CA; Manufactured by Varian Medical Systems Haan GmbH, Haan, Germany. To treat cancer of the uterus, cervix, endometrium and vagina.
- Product code
- JAQ System, Applicator, Radionuclide, Remote-Controlled
- Reason for recall
- Potential of weld failure of the Titanium intrauterine probe at the cervical stopper.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Varian Medical Systems sent an "URGENT MEDICAL DEVICE CORRECTION/URGENT FIELD SAFETY NOTICE" letter dated July 13, 2011 to all affected customers. The letter describes the product, problem, and the actions to be taken by the customers. The letter instructs customers to return all affected product to Varian Brachy Therapy by using the contact information provided in the letter. Replacement product will be provided free of charge to the customers. Customers should contact their local Varian Customer Support District or Regional Manager for questions regarding this notice.
- Quantity in commerce
- 437 - all sizes
- Distribution
- Worldwide Distribution.
- Product codes and lots
- Product number applicator Set: GM11006200;60 mm, 30 degree angle probe product number: GM11006160.