Z-2920-2011

Baxter Solution Set with Duo-Vent Spike; 109" (2.8 m), Male Luer Lock Adapter 10 drops/mL. Baxter Healthcare...

Baxter Healthcare Corporation / initiated 2011-07-21 / Terminated / root cause: manufacturing process

Baxter Healthcare Corp. began a recall of Baxter Solution Set with Duo-Vent Spike; 109" (2.8 m), Male Luer Lock Adapter 10 drops/mL. Baxter Healthcare Corporation. A sterile, single use, nonpyrogenic... on . The reason FDA records is "The Solution Sets are being recalled because the position of the slide clamp and roller clamp may have been reversed during assembly. The reversed orientation of the slide clamp and roller clamp will cause the set to be...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2920-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Baxter Healthcare Corporation
Product
Baxter Solution Set with Duo-Vent Spike; 109" (2.8 m), Male Luer Lock Adapter 10 drops/mL. Baxter Healthcare Corporation. A sterile, single use, nonpyrogenic fluid pathway for use with rigid nonvented and flexible containers.
Product code
FPA Set, Administration, Intravascular
Reason for recall
The Solution Sets are being recalled because the position of the slide clamp and roller clamp may have been reversed during assembly. The reversed orientation of the slide clamp and roller clamp will cause the set to be incorrectly loaded into an infusion pump, resulting in fluid flow from the patient to the pump, rather than from the pump to the patient.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
Baxter Healthcare Corporation sent a "Urgent Product Recall" letter dated July 21, 2011 to all affected customers. The letter included a description of the problem and requested they contact Baxter at 1-888-229-0001 to return all affected product. They were asked to complete and return the attached customer reply form, and to contact and notify their customers. For questions contact Baxter at 1-800-422-9837 during the hours of 8:00 am to 5:00 pm CST.
Quantity in commerce
4,752 units
Distribution
(USA) Nationwide Distribution including the states of Alabama, California, Iowa, Illinois, Indiana, Massachusetts, Maryland, Michigan, North Carolina, New Hampshire, Ohio, Pennsylvania, Tennessee and Virginia.
Product codes and lots
Product Code 1C8507s, lot numbers GR291542, GR292409, GR292797 and GR294454.