LATITUDE Patient Management System Communicator, Model 6482. The LATITUDE Patient Management system is...
Boston Scientific CRM Corp began a recall of LATITUDE Patient Management System Communicator, Model 6482. The LATITUDE Patient Management system is intended for use to remotely communicate with a... on . The reason FDA records is "Boston Scientific has identified two occurrences where the LATITUDE system did not process a data payload, which is the information uploaded to the LATITUDE system from a remote interrogation of the implanted device...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-3183-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Boston Scientific Corporation
- Product
- LATITUDE Patient Management System Communicator, Model 6482. The LATITUDE Patient Management system is intended for use to remotely communicate with a compatible Boston Scientific pulse generator and transfer data to a central database.
- Product code
- LWS Implantable Cardioverter Defibrillator (Non-Crt)
- Reason for recall
- Boston Scientific has identified two occurrences where the LATITUDE system did not process a data payload, which is the information uploaded to the LATITUDE system from a remote interrogation of the implanted device. These occurrences resulted in non-displayed payloads because the data is not posted to the website. If the payload is not displayed, the pertinent data will not be available on the
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Boston Scientific notified all affected customers by telephone on May 25, 2011. The customers were informed that there was an interrogation for one of their patients that was not able to be displayed on the website. Boston Scientific detected that a payload failure had occurred, but could not conclude what information was contained within the payload. Each patient has had previous and subsequent payloads that have processed and displayed on the LATITUDE website.
- Quantity in commerce
- 2
- Distribution
- Nationwide Distribution-in the states of Florida and Texas.
- Product codes and lots
- SN 035074 SN 032915