Lenstar LS 900 EyeSuite Biometry version i2.000, Four (4) Protective Cardboard Containers Per Each System...
Haag-Streit USA Inc began a recall of Lenstar LS 900 EyeSuite Biometry version i2.000, Four (4) Protective Cardboard Containers Per Each System. Haag-Streit USA, Inc. A non-invasive, non-contact... on . The reason FDA records is "A patient record may be linked to an incorrect/inconsistent set of calculation results under specific circumstances when using EyeSuite Software version i2.000". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2964-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Haag-Streit USA Inc
- Product
- Lenstar LS 900 EyeSuite Biometry version i2.000, Four (4) Protective Cardboard Containers Per Each System. Haag-Streit USA, Inc. A non-invasive, non-contact OLCR (Optical Low Coherence Reflectometer) Biometer used for obtaining ocular measurements and performing calculations to assist in determination of the appropriate power and type of IOL (Intraocular Lens) for implantation after removal of the natural crystalline lens following cataract removal.
- Product code
- HJB Instrument, Measuring, Corneal Radius
- Reason for recall
- A patient record may be linked to an incorrect/inconsistent set of calculation results under specific circumstances when using EyeSuite Software version i2.000.
- Root cause tag
- software
- FDA root cause
- Software change control
- Action
- Haag-Streit USA, Inc. sent an "Urgent Field Safety Notices" dated June 20, 2011 to all affected customers. The letter included a description of the affected product and problem. The firm instructed their customers to make sure all new and current templates are saved and upgrade to EyeSuite version i4.000. Customers are asked to complete and return the Software Update Verification Form. For additional information contact the Haag-Streit at (513) 398-3937.
- Quantity in commerce
- 216
- Distribution
- Nationwide (USA) Distribution.
- Product codes and lots
- Part Number 7220162. Serial Numbers: 21, 127, 223, 224, 284, 285, 316, 318, 319, 320, 322, 323, 324, 325, 326, 327, 328, 330, 331, 332, 333, 364, 365, 366, 368, 369, 371, 372, 376, 378, 379, 381, 384, 385, 386, 405, 407, 408, 409, 411, 412, 413, 414, 415, 416, 417, 418, 419, 420, 421, 423, 424, 443, 446, 447, 448, 449, 450, 452, 453, 454, 455, 456, 457, 458, 459, 460, 463, 469, 470, 481, 482, 483, 484, 485, 486, 487, 489, 490, 491, 492, 493, 494, 495, 496, 497, 499, 500, 527, 528, 529, 531, 532, 533, 535, 536, 537, 538, 539, 540, 541, 542, 543, 544, 545, 546, 547, 548, 549, 550, 551, 552, 558, 559, 560, 561, 563, 564, 565, 566, 567, 568, 569, 570, 571, 572, 573, 574, 576, 578, 579 581, 582, 583, 584, 585, 587, 588, 589, 590, 618, 619, 620, 621, 622, 624, 625, 626, 628, 630, 631, 632, 633, 634, 635, 636, 638, 639, 640, 641, 643, 644, 645, 647, 648, 649, 650, 651, 652, 653, 654, 656, 657, 658, 659, 660, 661, 662, 663, 665, 667, 668, 669, 670, 671, 672, 674, 675, 676, 677, 721, 723, 724, 725, 726, 727, 728, 729, 735, 824, 825, 922, 946, 966, 967, 968, 970, 1018, 1019, 1021, 1022, 1024, 1025, 1114, 1143 & 1195.