Z-2950-2011

CAPIOX Flexible Venous Reservoirs, 1500 mL, product codes: CXFVR1500X and CXFVRI500XC, used singly or...

Terumo Cardiovascular Systems Corporation / initiated 2011-07-21 / Terminated / root cause: manufacturing process

Terumo Cardiovascular Corporation began a recall of CAPIOX Flexible Venous Reservoirs, 1500 mL, product codes: CXFVR1500X and CXFVRI500XC, used singly or packaged as components in cardiovascular procedure trays... on . The reason FDA records is "Reservoir blood bag used in cardiovascular procedures may leak". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2950-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
CAPIOX Flexible Venous Reservoirs, 1500 mL, product codes: CXFVR1500X and CXFVRI500XC, used singly or packaged as components in cardiovascular procedure trays labeled in part ***Terumo Caerdiovascular Systems Corporation, Elkton, MD 21921*** Product Usage: The reservoir bag is intended for use as a blood collection and storage device during cardiopulmonary bypass procedures. The device is intended for use in conjunction with blood-gas oxygenators in the extra-corporeal circuit. The device may be used in procedures lasting up to 6 hours.
Product code
DTN Reservoir, Blood, Cardiopulmonary Bypass
Reason for recall
Reservoir blood bag used in cardiovascular procedures may leak.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Terumo Cardiovascular Systems Corporation sent an "URGENT MEDICAL DEVICE REMOVAL" letter dated July 20, 2011, to all affected consignee. The letter described the problem, the potential hazard, and the actions to be taken. The letter instructed customers to identify affected inventory and segregate from non-affected product. The recalling firm advises users to not use affected product and return for replacement. The customers are instructed to call the recalling firm Customer Service for a Returned Goods Authorization (RGA) number. Customers were asked to complete the attached Response Form and return as per the instructions. Consignees with questions should call 1-800-521-2818 Monday through Friday 8AM to 6PM EST.
Quantity in commerce
253 reservoirs
Distribution
(USA) Nationwide Distribution including states of: FL, MI, MT, NY, TN and VA.
Product codes and lots
CAPIOX Flexible Venous Reservoir, Lot NE13, (used as single units) or as a component in sterilized custom cardiovascular procedure kits with kit lot numbers NC21 R, ND21 R, NF02, and NF02.