Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-3081-2011

Mini MaxLock Extreme Plating System Case & Tray, Model Number(s): MXM-901

TORNIER Inc. / initiated 2011-07-06 / Terminated / root cause: manufacturing process

Orthohelix Surgical Designs Inc began a recall of Mini MaxLock Extreme Plating System Case & Tray, Model Number(s): MXM-901, Product Usage: The Mini MaxLock Extreme Plating System is intended to stabilize and... on . The reason FDA records is "The surgical tray does not meet our Instruction for Use (IFU) for dry time requirement". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-3081-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
TORNIER Inc.
Product
Mini MaxLock Extreme Plating System Case & Tray, Model Number(s): MXM-901, Product Usage: The Mini MaxLock Extreme Plating System is intended to stabilize and aid in the repair of fractures, fusions, and osteotomies for small bones and bone fragments. Surgical Case and Tray Delivery System used to house plating and screw orthopedic implant system and instrumentation.
Product code
HRS Plate, Fixation, Bone
Reason for recall
The surgical tray does not meet our Instruction for Use (IFU) for dry time requirement.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Orthohelix Surgical Designs Inc sent a RECALL NOTIFICATION letter dated July 6, 2011 and an UPDATED URGENT RECALL 2ND NOTIFICATION letter dated July 13, 2011 to all affected customers. The letter identified the recalled product, the problem, and the Field Correction actions to be taken. Customers were instructed to remove the blue rubber mat in the utility bay of the tray base immediately, tag the mat with the tray number, and return it to recalling firm immediately. The letter states once customers have removed the blue rubber mat from the tray base, they may continue to use the tray and system normally. No further actions or precautions are needed. Distributors were also asked to notify their customers of the recall. If you have any questions, please call 330-247-1444 or 330-247-1445.
Quantity in commerce
61 trays
Distribution
USA Nationwide Distribution including the states of: AR, AZ, CA, CO, FL, GA, IL, IN, KY, MA, MD, MI, MO, NC, NE, NJ, NM, NY, OH, OR, PA, SD, TN, TX, UT, VA, WA, WI & WY.
Product codes and lots
Model Number(s): MXM-901; Tray codes affected: LAQ, LAV, LCL, LCN, LBZ, LCA, LBV, LCU, LAR, LBJ, LBT, LDE, LDG, LBL, LCD, LBI, LCI, LBQ, LCV, LBK, LBC, LCW, LCB, LDH, LBM, LCM, LCK, LDB, LBS, LCE, LDD, LDF, LAX, LDI, LAP, LBH, LBP, LAO, LBB, LBE, LBX, LCH, LAT, LAS, LCR, LDC, LAW, LCJ, LDJ, LCZ, LAU, LBF, LCQ, LBN, LCP, LCX, LBU, LBW, LBY, LCT, LCY, LAY, LAZ, LBA, LBD, LBG, LBO, LBR, LCC, LCF, LCG, LCS, LDA, LDM, LDN, LDO, LDP, LDQ, LDR, LDS, LDT, LDU & LDK.