Contour, Contour TS Blood Glucose Test Strips and Bonus Packs Intended for self-testing by people with...
Bayer Healthcare, LLC began a recall of Contour, Contour TS Blood Glucose Test Strips and Bonus Packs Intended for self-testing by people with diabetes and healthcare professionals to monitor glucose... on . The reason FDA records is "All 10- or 25- count glucose test strip vials, either loose vials or found in patient starter kits of Contour, Contour TS and Contour" USB blood glucose meters are being recalled a packaging issue which, under certain...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-3268-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medrad, Inc.
- Product
- Contour, Contour TS Blood Glucose Test Strips and Bonus Packs Intended for self-testing by people with diabetes and healthcare professionals to monitor glucose concentrations in whole blood
- Product code
- LFR Glucose Dehydrogenase, Glucose
- Reason for recall
- All 10- or 25- count glucose test strip vials, either loose vials or found in patient starter kits of Contour, Contour TS and Contour" USB blood glucose meters are being recalled a packaging issue which, under certain conditions, can cause some test strips packaged in some small-count vials to operate outside of US FDA-cleared performance specifications and produce a low-bias reading during blood
- Root cause tag
- packaging
- FDA root cause
- Packaging
- Action
- Bayer sent an "URGENT DIABETES TEST STRIP RECALL" letter dated August 12, 2011, to its customers with the name of the test strips, sizes recalled, reason for the recall, potential injury, and instructions to discontinue use and return the test strips to Bayer. The recall letter also included a recall response form that was to be faxed back to Bayer at 1-800-876-2243. For questions call Customer Logistics at 1-800-348-2637.
- Quantity in commerce
- 3,617,287 vials/ 54,555,460 test strips
- Distribution
- Nationwide Distribution and Puerto Rico
- Product codes and lots
- 9545C, 9677, 9678, 9679, 9680, 7151G, 7190, 7183, 9615, 7393, 9591, 9591A 9579, 1802, 6181, 7070A, 9725, 9507C, and 9578. Contour 125 Count Bonus Pack (includes 2-50 Count and 1-25 Count Vials) the following lot numbers:DW0AC3B32A, DW0AC3B32C, DW0BC3B31A, DW0BC3B31B, DW0BC3D05B, DW0CC3B31A, DW0CC3B31B, DW0GC3B31C, DW0GC3B31D, DW0GC3B31F, DW0HC3B32A, DW9MC3B32D, DW9MC3B32E, DW9MC3B33A, and DW9MC3B33B.