Z-3038-2011

Duo-Decapolar Electrophysiology Catheter, Manufacturing Part Number D-1171-35-S, Catalog Number D728260RT...

Biosense Webster Inc / initiated 2011-04-29 / Terminated / root cause: labeling and UDI

Biosense Webster, Inc. began a recall of Duo-Decapolar Electrophysiology Catheter, Manufacturing Part Number D-1171-35-S, Catalog Number D728260RT Duo-Decapolar Electrophysiology Catheter has been... on . The reason FDA records is "The recall was expanded because between February and April 2011, Biosense Webster, Inc. notified customers that a voluntary Customer Notification had been Initiated due to the inaccessibility of the electronic version...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-3038-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Biosense Webster Inc
Product
Duo-Decapolar Electrophysiology Catheter, Manufacturing Part Number D-1171-35-S, Catalog Number D728260RT Duo-Decapolar Electrophysiology Catheter has been designed for electrophysiological mapping of cardiac structures.
Product code
DRF Catheter, Electrode Recording, Or Probe, Electrode Recording
Reason for recall
The recall was expanded because between February and April 2011, Biosense Webster, Inc. notified customers that a voluntary Customer Notification had been Initiated due to the inaccessibility of the electronic version of the Instructions for Use (e-IFU) for the "10 ft. Interface Cable, 10 Pin - Shielded Tip Pins to 12 Pin - Blue", Catalog Number 0128709, through the J&J Gateway (i.e., e-IFU websi
Root cause tag
labeling and UDI
FDA root cause
Labeling Change Control
Action
Biosense Webster sent an "URGENT FIELD SAFETY NOTICE" letter dated April 29, 2011 to all affected customers. The letter describes the product, problem, and the action taken by the firm. The letter notifies and advises customers that the updated IFUs for ISMUS Catheter with Auto ID Technology and the Duo-Decapolar catheters have been released and posted on the e-IFU website as of April 22, 2011 and April 21, 2011. Customers were instructed to complete the Acknowledgement Form and return the form to Biosense. Customers with questions or require additionial information were instructed to contact their Biosense Webster representative.
Quantity in commerce
1530 units
Distribution
Nationwide Distribution
Product codes and lots
Not available