Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0303-2012

Battery Module for Sarns Modular Perfusion System 8000

Terumo Cardiovascular Systems Corporation / initiated 2011-10-21 / Terminated / root cause: manufacturing process

Terumo Cardiovascular Systems Corporation began a recall of Battery Module for Sarns Modular Perfusion System 8000 Product Usage: The Sarns Modular Perfusion System 8000 is indicated for up to 6 hours use in... on . The reason FDA records is "Terumo Cardiovascular Systems is aware on one instance in which an incorrectly labeled printed circuit board for the Sarns Modular Perfusion System 8000 was shipped to a single customer. The affected board mislabeled...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0303-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
Battery Module for Sarns Modular Perfusion System 8000 Product Usage: The Sarns Modular Perfusion System 8000 is indicated for up to 6 hours use in extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures only when used by a qualified perfusionist.
Product code
DWB Pump, Blood, Cardiopulmonary Bypass, Roller Type
Reason for recall
Terumo Cardiovascular Systems is aware on one instance in which an incorrectly labeled printed circuit board for the Sarns Modular Perfusion System 8000 was shipped to a single customer. The affected board mislabeled two connection between the board and the transformer.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
Terumo conducted a Correction/Removal - Customer Communicated Script on November 7, 2011 to the affected customer. The phone script communicated the correction and reason for removal, potential hazards, affected population, and confirmation of possession of affected module. A replacement date was set for November 7, 2011. For questions or concerns contact Terumo CVS Customer Service at 1-800-521-2818, hours are Monday to Friday. 8 AM -6 PM EST.
Quantity in commerce
10 units
Distribution
Nationwide Distribution (USA) including the state of San Antonio, TX
Product codes and lots
Catalog Number 16422 Serial Number 1901