Battery Module for Sarns Modular Perfusion System 8000
Terumo Cardiovascular Systems Corporation began a recall of Battery Module for Sarns Modular Perfusion System 8000 Product Usage: The Sarns Modular Perfusion System 8000 is indicated for up to 6 hours use in... on . The reason FDA records is "Terumo Cardiovascular Systems is aware on one instance in which an incorrectly labeled printed circuit board for the Sarns Modular Perfusion System 8000 was shipped to a single customer. The affected board mislabeled...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0303-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Terumo Cardiovascular Systems Corporation
- Product
- Battery Module for Sarns Modular Perfusion System 8000 Product Usage: The Sarns Modular Perfusion System 8000 is indicated for up to 6 hours use in extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures only when used by a qualified perfusionist.
- Product code
- DWB Pump, Blood, Cardiopulmonary Bypass, Roller Type
- Reason for recall
- Terumo Cardiovascular Systems is aware on one instance in which an incorrectly labeled printed circuit board for the Sarns Modular Perfusion System 8000 was shipped to a single customer. The affected board mislabeled two connection between the board and the transformer.
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- Terumo conducted a Correction/Removal - Customer Communicated Script on November 7, 2011 to the affected customer. The phone script communicated the correction and reason for removal, potential hazards, affected population, and confirmation of possession of affected module. A replacement date was set for November 7, 2011. For questions or concerns contact Terumo CVS Customer Service at 1-800-521-2818, hours are Monday to Friday. 8 AM -6 PM EST.
- Quantity in commerce
- 10 units
- Distribution
- Nationwide Distribution (USA) including the state of San Antonio, TX
- Product codes and lots
- Catalog Number 16422 Serial Number 1901