Z-0056-2012
8F Smart Port CT, Catalog number: CT80STPD and CT80STPD-NF; AngioDynamics, INC. The Smart Port CT Implantable...
Boston Scientific Corporation / initiated 2011-08-01 / Terminated / root cause: design
Angiodynamics, INC began a recall of 8F Smart Port CT, Catalog number: CT80STPD and CT80STPD-NF; AngioDynamics, INC. The Smart Port CT Implantable Port is a Single Titanium Port System with... on . The reason FDA records is "The product may contain an incorrect size catheter and introducer". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0056-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Boston Scientific Corporation
- Product
- 8F Smart Port CT, Catalog number: CT80STPD and CT80STPD-NF; AngioDynamics, INC. The Smart Port CT Implantable Port is a Single Titanium Port System with Attachable 8.0 Fr. Polyurethane Catheters and 8Fr. Introducer Kit.
- Product code
- DYB Introducer, Catheter
- Reason for recall
- The product may contain an incorrect size catheter and introducer.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- AngioDynamics, Inc contacted affected customers initially via telephone and a follow up "Notification of Medical Device Correction" letter was sent dated August 1, 2011. The letter informed them of the problem and provided instructions. Customers were instructed to utilize the 8F Introducer Sheath provided within the 8F Smart Port CT Implantable Port during port implantation. The 8F Valved Sheath (V18) may continue to be utilized during the placement of the 7.5F Smart Port Implantable Port (for example: CT75STSD; CT75STSD-NF). For question on this recall contact AngioDynamics, Inc at (800) 772-6446.
- Quantity in commerce
- 1095 total units
- Distribution
- (USA) Nationwide distribution including the states of CA, FL, GA, IA, IL, IN, KY, MD, MI, MO, NC, NE, NJ, NY, OH, OK, OR, TN, UT, WA and WI.
- Product codes and lots
- Catalog number: CT80STPD and CT80STPD-NF