Z-3214-2011
Sarns 8000 Cardioplegia Monitor, Catalog Number 16414; also known as PROG COMP BD AMPRO 420 CPG 8K, Catalog...
Terumo Cardiovascular Systems Corporation / initiated 2011-08-29 / Terminated
Terumo Cardiovascular Systems Corporation began a recall of Sarns 8000 Cardioplegia Monitor, Catalog Number 16414; also known as PROG COMP BD AMPRO 420 CPG 8K, Catalog Number 816373 The Sarns 8000 Cardioplegia Monitor... on . The reason FDA records is "The Cardioplegia Monitor for the Sarns Modular Perfusion System 8000 may fail to track cardioplegia delivery volume". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-3214-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Terumo Cardiovascular Systems Corporation
- Product
- Sarns 8000 Cardioplegia Monitor, Catalog Number 16414; also known as PROG COMP BD AMPRO 420 CPG 8K, Catalog Number 816373 The Sarns 8000 Cardioplegia Monitor is indicated to measure extracorporeal line pressure and signal when the pressure exceeds the operator-set limit.
- Product code
- DWB Pump, Blood, Cardiopulmonary Bypass, Roller Type
- Reason for recall
- The Cardioplegia Monitor for the Sarns Modular Perfusion System 8000 may fail to track cardioplegia delivery volume.
- FDA root cause
- Component change control
- Action
- Terumo Cardiovascular Systems sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated August 29, 2011 to all affected customers. The letter describes the product, problem, and action to be taken. The letter advises customers to make sure all users are aware of the notice. Customers are instructed to complete and return an attached Customer Response Form. Terumo will replace the software on all monitors of the affected product. Customers with questions or concerns should call Terumo CVS Customer Service 1-800-521-2818.
- Quantity in commerce
- 85
- Distribution
- Worldwide Distribution--USA (nationwide) including the states of AL, AR, CA, CO, GA, IL, IN, MS, NC, OH, TN, VA, and WI and countries of Belgium, Columbia, Hong Kong, India, Singapore, and Taiwan.
- Product codes and lots
- Cardioplegia Catalog Number 16414 and Serial Numbers 2060, 3305, 3763, 3823, 3834, 3835, 3851-3876, 3878-3899, 3901, 3902, 3904, 3906-3913, 3915-3933 and 3935. Also, catalog number 816373, Serial numbers 491212, 498329, 505891, 505892, 509307, 510221, 512688, 514493, 517238, 522537, 524438, 527767, 534182, 541101, 545173, 551503, 557425, 563225, 568962 and 576045.