Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-3067-2011

Applied Medical 15x100 mm Kii Optical Access Systems, Non-Threaded 6/Box, Model C0R37

Applied Medical Resources Corp / initiated 2009-12-18 / Terminated / root cause: design

Applied Medical Resources Corp began a recall of Applied Medical 15x100 mm Kii Optical Access Systems, Non-Threaded 6/Box, Model C0R37 Product Usage: Applied Medical's C0R36, C0R37, C0R38, and C0R39 trocars... on . The reason FDA records is "The recall was expanded on December 22, 2009 to include four models: C0R36, C0R37, C0R38, and C0R39. Initially, Applied Medical conducted a voluntary recall on June 9, 2009 of two lot numbers, 1084426 and 1085506, of...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-3067-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Applied Medical Resources Corp
Product
Applied Medical 15x100 mm Kii Optical Access Systems, Non-Threaded 6/Box, Model C0R37 Product Usage: Applied Medical's C0R36, C0R37, C0R38, and C0R39 trocars are diposable single-use devices indicated for use in general, abdominal, gynecological and thoraic minimally invasive surgeries. Trocars establish a path of entry through tissue planes and into internal spaces for purposes of inserting endoscopic instruments that are used to perform surgery on internal organs.
Product code
GCJ Laparoscope, General & Plastic Surgery
Reason for recall
The recall was expanded on December 22, 2009 to include four models: C0R36, C0R37, C0R38, and C0R39. Initially, Applied Medical conducted a voluntary recall on June 9, 2009 of two lot numbers, 1084426 and 1085506, of the 15 mm Kii non-threaded optical access system, model C0R36, due to a potential weakness in the cannula tip, possibly resulting in a tip breakage.
Root cause tag
design
FDA root cause
Component design/selection
Action
Applied Medical Resources Corp. notified customers via an overnight "URGENT: MEDICAL DEVICE RECALL" letter dated December 22, 2009. The letter informed customers of the recalled products, problem, and actions to be taken. The letter asked that the listed recalled products be returned immediately. The customers were asked to check their inventory, complete the attached Voluntary Recall Notification Confirmation Form and fax as per the instructions. The recall letter provided customers with a contact and telephone number for product return questions (1-949-713-8652) and a separate contact for regulatory questions (1-949-713-8369).
Quantity in commerce
13,791 units total
Distribution
Worldwide Distribution - Nationwide in the USA and to countries of: Australia, Austria, Belgium, Canada, Chile, Columbia, Cyprus, Denmark, Finland, France, Germany, Great Britain, Hong Kong, India, Ireland, Italy, Malaysia, Netherlands, New Zealand, Norway, Poland, Portugal, Saudi Arabia, Singapore, South Africa, and Switzerland.
Product codes and lots
1077612, 1078134, 1079448, 1081556, 1082011, 1083371, 1084938, 1085509, 1087279, 1089638, 1088096, 1090841, 1091538, 1093144, 1097420