Terumo Advanced Perfusion System 1 Base, 100/120V, Catalog 801763, 6 " Roller Pump Occlusion: Reference...
Terumo Cardiovascular Systems Corporation began a recall of Terumo Advanced Perfusion System 1 Base, 100/120V, Catalog 801763, 6 " Roller Pump Occlusion: Reference AA-2011-015 C. Indicated for use for up to 6 hours on... on . The reason FDA records is "Terumo CVS has received reports of 27 instances in which the 6 inch Roller Pump jammed during cardioplegia delivery. There has been one report of the pump jamming during a case that concluded with an adverse patient...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1320-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Terumo Cardiovascular Systems Corporation
- Product
- Terumo Advanced Perfusion System 1 Base, 100/120V, Catalog 801763, 6 " Roller Pump Occlusion: Reference AA-2011-015 C. Indicated for use for up to 6 hours on the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures.
- Product code
- DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass
- Reason for recall
- Terumo CVS has received reports of 27 instances in which the 6 inch Roller Pump jammed during cardioplegia delivery. There has been one report of the pump jamming during a case that concluded with an adverse patient outcome, although there was no indication that the pump jam contributed to the adverse patient outcome.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- "URGENT MEDICAL DEVICE RECALL" letters were sent on 3/16/12 to all consignees detailing the correction as well as also adding an Addendum to the Operator's manual for the Terumo System 1 to incorporate additional information and warnings for use when operating a 6" roller pump with dual-sized tubing. For questions call 1-800-521-2818.
- Quantity in commerce
- 994 units
- Distribution
- Worldwide Distribution--USA, including the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, WV, Puerto Rico and the countries of INDIA, CANADA, CHILE, HONG KONG, EGYPT, PAKISTAN, KUWAIT, SAUDI ARABIA, JORDAN, UNITED ARAB EMIRATES, SOUTH AFRICA, SOUTH KOREA, PHILIPPINES, JAPAN, PHILIPPINES, SINGAPORE, AUSTRALIA, MALAYSIA, TAIWAN, RUSSIA, TURKEY, BELGIUM, GERMANY, COLUMBIA, COSTA RICA, ARGENTINA, HONDURAS, COLUMBIA, DOMINICAN REPUBLIC, GUATEMALA, MEXICO, VIETNAM, and THAILAND.
- Product codes and lots
- Lot numbers: 0011-0610,1001-1034,1037-1049,1100,1103-1445, and 1447-1449.