Entroy Pool Lift; a patient lift used in conjunction with a chair or stretcher to lift them in/out of a...
Arjo, Inc. dba ArjoHuntleigh began a recall of Entroy Pool Lift; a patient lift used in conjunction with a chair or stretcher to lift them in/out of a swimming/therapy pool; Arjo Hospital Equipment AB... on . The reason FDA records is "The Entroy Pool Lift boom may rapidly lower un-commanded to the bottom of its stroke, with a potential for injury to the resident or caregiver". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-3070-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Arjo, Inc.
- Product
- Entroy Pool Lift; a patient lift used in conjunction with a chair or stretcher to lift them in/out of a swimming/therapy pool; Arjo Hospital Equipment AB, Eslov, Sweden; Models GAB1000-01-xx, GAB1000-11-xx and GAC100x Entroy is a pool lift that docks to either a chair or a stretcher and aids disabled people getting in and out of a pool. This equipment is intended for disabled persons who need to transfer in and out of a therapy and swimming pool under supervision of skilled and trained personnel in accordance with the Entroy Operating and Product Care Instructions. All other use must be avoided. Active/semi-active persons can and should use the Entroy chair (with or without chassis) while bedridden persons without trunk stability can and should be transferred with the Entroy stretcher with chassis.
- Product code
- FSA Lift, Patient, Non-Ac-Powered
- Reason for recall
- The Entroy Pool Lift boom may rapidly lower un-commanded to the bottom of its stroke, with a potential for injury to the resident or caregiver.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- ArjoHuntleigh sent an Urgent Field Correction Notice I.D. 110701AHE to the affected end user via UPS 2nd Day on August 4, 2011. The notice informed the account of the potential for lift boom to rapidly lower un-commanded to the bottom of its stroke. The account was requested to immediately cease use of the affected lift, and disconnect the power to the lift to prevent further use until corrected. The account was requested to complete and sign the Customer Response Form, and return it to ArjoHuntleigh to schedule a service call to correct the lift. Questions were directed to the ArjoHuntleigh at 1-800-323-1245, ext 54485.
- Quantity in commerce
- 1 unit in the U.S., 311 units in the rest of the world
- Distribution
- Worldwide Distribution - USA including CT
- Product codes and lots
- Model GAB1000-01-US, serial number SEE0326028 Rest of the world: Models GAB1000-01-xx, GAB1000-11-xx and GAC100x, all serial numbers
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
