Z-3129-2011

VIDAS Chlamydia test kits VIDAS Chlamydia is an automated qualitative test for use on the VIDAS family...

bioMerieux, Inc. / initiated 2011-07-08 / Terminated

bioMerieux, Inc. began a recall of VIDAS Chlamydia test kits VIDAS Chlamydia is an automated qualitative test for use on the VIDAS family instruments, for the detection of Chlamydia antigen in... on . The reason FDA records is "Complaints were received of high rates of false positive results, leading to unnecessary antibiotic therapy being prescribed to patients". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-3129-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
bioMerieux, Inc.
Product
VIDAS Chlamydia test kits VIDAS Chlamydia is an automated qualitative test for use on the VIDAS family instruments, for the detection of Chlamydia antigen in endocervical or urethral specimens, using the ELFA technique (Enzyme Linked Fluorescent Assay.
Product code
LJC Enzyme Linked Immunoabsorbent Assay, (Chlamydiae Group)
Reason for recall
Complaints were received of high rates of false positive results, leading to unnecessary antibiotic therapy being prescribed to patients.
FDA root cause
Other
Action
bioMerieux sent an "URGENT PRODUCT REMOVAL NOTICE" letter dated July 8, 2011 to all affected customers. The letter describes the product, problem, and the actions to be taken by the customers. The letter instructs customers to discontinue using and to destroy any of the affected product. The firm advises customers to confirm all positive results obtained with the affected lot number with another method prior to reporting results. An Acknowledgement and Product Replacement form was attached for customers to complete and return to the firm. Customers should contact their local bioMerieux Customer Service representative for questions regarding this notice.
Quantity in commerce
283 units
Distribution
Nationwide Distribution- including CA, DC, FL, IL, IL, IN, MD, MI, MN, MO, MT, NC, NE, NJ, NY, PA, UT, VA and WI.
Product codes and lots
Catalog number: 30101-01, Lot number 868400501.