Z-3159-2011

VARIANT II TURBO Link Hemoglobin A1c Program Reorder Pack. Calibrator/Diluent Set. IVD. Kit components...

Bio-Rad Laboratories Inc. / initiated 2011-07-28 / Terminated / root cause: manufacturing process

Bio-Rad Laboratories, Inc. began a recall of VARIANT II TURBO Link Hemoglobin A1c Program Reorder Pack. Calibrator/Diluent Set. IVD. Kit components: Elution Buffer A, 2 x 2500 mL Elution Buffer B, 1 x... on . The reason FDA records is "The Elution Buffer B, lot no. GA 10430, was shipped in a 2.0 L bottle instead of the standard 2.5 L bottle". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-3159-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Bio-Rad Laboratories Inc.
Product
VARIANT II TURBO Link Hemoglobin A1c Program Reorder Pack. Calibrator/Diluent Set. IVD. Kit components: Elution Buffer A, 2 x 2500 mL Elution Buffer B, 1 x 2100 mL Wash/Diluent Solution, 1 x 2500 mL Cartridge Set, 2 Cation exchange Analytical Cartridges, 4 Cation Guard Cartridges; CD-ROM - Calibration/Diluent Set, 2 levels, 2 vials each of lyophilized human red blood cell hemolysate with preservative, 1 bottle of diluent, 100 mL, deionized water Whole Blood Primer, 6 x 1 mL Sample Vials, 2 x 100 Instruction Manual. Bio-Rad Laboratories Inc. Hercules, CA 94547 Intended for the percent determination of hemoglobin A1c in human whole blood using ion-exchange high-performance liquid chromatography (H LPC).
Product code
LCP Assay, Glycosylated Hemoglobin
Reason for recall
The Elution Buffer B, lot no. GA 10430, was shipped in a 2.0 L bottle instead of the standard 2.5 L bottle.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
The firm, BIO-RAD, sent a "Medical Device Correction" letter dated July 28, 2011 to all consignees/customers via Fed Ex on August 8, 2011. The letter described the product, problem and actions to be taken. The customers were informed that they could continue to use and report results with the affected buffer lot and the stability studies will continue until lot expiration date. The consignees were instructed to distribute this letter to all customers that have received the affected lot and to distribute the MEDICAL DEVICE CORRECTION RESPONSE FORM only to those customers that request replacement Elution Buffer B. In addition, the customers were to complete and return the MEDICAL DEVICE CORRECTION RESPONSE FORM via fax to (510) 741-3954, Attention: BIO-Rad CSD Regulatory Affairs Department. If you have any questions, please contact your regional Bio-Rad office or 510-724-7000.
Quantity in commerce
40
Distribution
Nationwide distribution: MD, MI, MN, NY, OH, SC, VA and WI.
Product codes and lots
Catalog Number: 270-2716; Reorder Pack Lot No: 70210702; Expires September 30, 2011. Elution Buffer B: Lot number 270-2718, lot number: GA10430, Exp 2/29/2012.