Bausch & Lomb Stellaris Optimized Stability Vacuum Pack, Ref: BL5112, packaged 6 units/case, RX, Sterile...

Bausch & Lomb Incorporated / initiated 2011-07-27 / Terminated / root cause: manufacturing process

Bausch & Lomb Inc began a recall of Bausch & Lomb Stellaris Optimized Stability Vacuum Pack, Ref: BL5112, packaged 6 units/case, RX, Sterile. Bausch & Lomb Incorporated, Rochester, NY For use... on . The reason FDA records is "The aspiration channel on the affected vacuum cassettes may be blocked due to a molding malfunction". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-3089-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Bausch & Lomb Incorporated
Product
Bausch & Lomb Stellaris Optimized Stability Vacuum Pack, Ref: BL5112, packaged 6 units/case, RX, Sterile. Bausch & Lomb Incorporated, Rochester, NY For use with the Bausch & Lomb Stellaris Vision Enhancement System for the phacoemulsification of an opacified crystalline lens during anterior segment surgery
Product code
HQC Unit, Phacofragmentation
Reason for recall
The aspiration channel on the affected vacuum cassettes may be blocked due to a molding malfunction
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Bausch & Lomb Inc. notified all affected customers by direct telephone contact on July 27, 2011 and a follow-up FAX or e-mail dated July 27, 2011 The letter included; affected product, problem, and requests customers return the product. An acknowledgement form was included to complete and return with the product. For question on this recall contact Bausch & Lomb at (949) 521-7895.
Quantity in commerce
2/6-unit cases
Distribution
Nationwide Distribution.
Product codes and lots
Model # BL5112, Lot #U6476,  Exp. 11/2012