Verify SixCess Challenge Pack, Model #'s: LCC003, LCC005, LCC014 & LCC310, Mfr. By: Albert Browne, LTD...
Steris Corporation began a recall of Verify SixCess Challenge Pack, Model #'s: LCC003, LCC005, LCC014 & LCC310, Mfr. By: Albert Browne, LTD, Chancery House 190 Waterside Rd. Hamilton Industrial... on . The reason FDA records is "During an inspection the firm was notified that language contained in labeling and promotional literature directs Customers that the products can be used to replace biological indicators and/or release loads containing...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-3046-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- STERIS Corporation
- Product
- Verify SixCess Challenge Pack, Model #'s: LCC003, LCC005, LCC014 & LCC310, Mfr. By: Albert Browne, LTD, Chancery House 190 Waterside Rd. Hamilton Industrial Park, Leicester, Leicester, LE5QZ, United Kingdom. Verify SixCess Challenge Packs are used to confirm that that critical parameters of steam sterilization have been reached within a challenging test pack design. Each challenge pack is also specific to a sterilization temperature and type (i.e. 275F 3 minute Prevacuum Cycle).
- Product code
- JOJ Indicator, Physical/Chemical Sterilization Process
- Reason for recall
- During an inspection the firm was notified that language contained in labeling and promotional literature directs Customers that the products can be used to replace biological indicators and/or release loads containing implants.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling False and Misleading
- Action
- STERIS sent a Customer Notification letter dated March 16, 2010, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to review and follow the Challenge Packs' revised instructions for use enclosed as Attachment A (270-4) and Attachment B (275F 3). For any questions regarding this recall customers were instructed to contact their STERIS Account Manager or call 440-392-7421.
- Quantity in commerce
- 2,500
- Distribution
- Worldwide Distribution - USA (nationwide) and the countries of Lebanon and Qatar.
- Product codes and lots
- Model #'s: LCC003, LCC005, LCC014 & LCC310.
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
