Comprehensive Reverse Humeral Fracture Inserter REF 406993 Comprehensive Instrumentation Reverse Humeral...
Biomet, Inc. began a recall of Comprehensive Reverse Humeral Fracture Inserter REF 406993 Comprehensive Instrumentation Reverse Humeral Inserter QTY 1. Instrument that is attached to the... on . The reason FDA records is "Recall initiated in response to a Biomet Canada report, which noted that while testing the instruments, the pin used to secure the implant on the device did not engage on the effected lots and were distributed". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-3289-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- Comprehensive Reverse Humeral Fracture Inserter REF 406993 Comprehensive Instrumentation Reverse Humeral Inserter QTY 1. Instrument that is attached to the trial prosthesis for placement in the prepared bone canal to check for correct sizing, version and humeral length. Also used to insert the humeral implant into the canal.
- Product code
- LXH Orthopedic Manual Surgical Instrument
- Reason for recall
- Recall initiated in response to a Biomet Canada report, which noted that while testing the instruments, the pin used to secure the implant on the device did not engage on the effected lots and were distributed.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- On 8/3/11 Biomet Orthopedics sent an "URGENT MEDICAL DEVICE RECALL NOTICE" letter to all affected customers. The letter included: reason for the recall, risks associated with use of the device, and alternate instructions for use of the recalled device. Customers are instructed to locate the affected products, discontinue use, return the devices, and a fax a copy of the response form. For information or questions on this recall call Biomet at (574) 372-1570.
- Quantity in commerce
- 6 units (internationally)
- Distribution
- International Distribution Only: Canada, Australia, and Columbia.
- Product codes and lots
- Part numbers: 406993, Lot numbers:176070 and 657550; and 406993-00, Lot number : 203119.
FIRST ARTICLE / WEEKLY