Cardiovascular Custom Procedure Kits, REV: 2A, REV: 2B X-COATED FX PK Item Number: 73191 Terumo...
Terumo Cardiovascular Systems Corp began a recall of Cardiovascular Custom Procedure Kits, REV: 2A, REV: 2B X-COATED FX PK Item Number: 73191 Terumo Cardiovascular Systems, Ashland, MA 01721 The Cardiovascular... on . The reason FDA records is "3-T Medical Systems one way vent valves, included in the Custom Terumo Cardiovascular kits, may separate at the bond that holds the two halves". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-3247-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Terumo Cardiovascular Systems Corporation
- Product
- Cardiovascular Custom Procedure Kits, REV: 2A, REV: 2B X-COATED FX PK Item Number: 73191 Terumo Cardiovascular Systems, Ashland, MA 01721 The Cardiovascular Procedure Kits are indicated for use only in the extracorporeal circuit for the cardiopulmonary bypass procedure(s) for which the user designed it. The 3-T L. V. Control Valve is a one-way vent valve used during cardiopulmonary bypass surgery to help prevent the buildup of excess vacuum pressure when suctioning either from the heart or the surgical field, and to help prevent inadvertent retrograde now and air into the heart.
- Product code
- MJJ Cpb Check Valve, Retrograde Flow, In-Line
- Reason for recall
- 3-T Medical Systems one way vent valves, included in the Custom Terumo Cardiovascular kits, may separate at the bond that holds the two halves
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Terumo Cardiovascular Systems initiated contact on August 12, 2011 by phone, fax, email and Urgent: Medical Device Recall mail to all affected customers. The firm advised customers of the affected product, problem and required recall actions. Customers can remove whole customized packs that are not in urgent medical need and return packs. Or the User may opt to use a sterile LH 130 vent valve from Terumo Cardiovascular Systems. A response form was included to return back to the firm. For questions on this recall contact Terumo Customer Service at (734) 525-9580.
- Quantity in commerce
- 160 units
- Distribution
- Nationwide Distribution, including the states of FL and MI.
- Product codes and lots
- Lot Numbers: MM11 MN15 MP20