Z-3092-2011
Evident MWA Generator, REF VTGEN, Output: 60 W / 50 Ohm, Valleylab, a division of Tyco Healthcare Group LP...
Medtronic, Inc. / initiated 2009-02-06 / Terminated / root cause: design
Covidien, PLC began a recall of Evident MWA Generator, REF VTGEN, Output: 60 W / 50 Ohm, Valleylab, a division of Tyco Healthcare Group LP Boulder, CO 80301. Intended for the coagulation of... on . The reason FDA records is "Generator may deliver full power when a lower power is selected". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-3092-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic, Inc.
- Product
- Evident MWA Generator, REF VTGEN, Output: 60 W / 50 Ohm, Valleylab, a division of Tyco Healthcare Group LP Boulder, CO 80301. Intended for the coagulation of soft tissue. Not intended for use in cardiac procedures.
- Product code
- GEI Electrosurgical, Cutting & Coagulation & Accessories
- Reason for recall
- Generator may deliver full power when a lower power is selected.
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- Covidien sales reps began notifying affected customers on February 6, 2009, followed by a letter. The letter identified the product, the problem, and the action to be taken by the customer. Customers were asked to return the affected product to: Covidien Energy Based Devices 5920 Longbow Road Boulder, Colorado 80301 Customers were asked to label the exterior of the carton with their Return Authorization number. Covidien EbD would address the issue and return the unit to them. For any questions customers were asked to call Covidien EbD's Service Department toll-free at 1-800-255-8522, x 4923 or contact their local sales representative for additional information.
- Quantity in commerce
- 357 units
- Distribution
- Worldwide Distribution - USA (nationwide) and the countries of Belgium, Denmark, France, Germany, Hong Kong, Hungary, Israel, Italy, Netherlands, Poland, Portugal, Romania, Singapore, Slovakia, Spain, Switzerland, and the UK.
- Product codes and lots
- Serial Numbers: XP7K0101R to XP7K0120R, XP7L0121R to XP7L0140R, XP8B0141R to XP8B0167R, XP8C0168R to XP8C0231R, XP8D0232R to XP8D0320R, XP8E0321R to XP8E0390R, XP8F0391R to XP8F0485R, XP8H0578R to XP8H0615R, XP8I0626R to XP8I0660R, XP8J0661R to XP8J0690R, XP8L0701R to XP8L0730R.