Z-0034-2012

Axle Interspinous Fusion System Torque Driver Handle, Model #: X060-0323, Lot # 525941, S/Ns 001-017, Lot #...

Xtant Medical Holdings, Inc. / initiated 2011-07-15 / Terminated / root cause: design

X Spine Systems Inc began a recall of Axle Interspinous Fusion System Torque Driver Handle, Model #: X060-0323, Lot # 525941, S/Ns 001-017, Lot # 525942, S/Ns 001-050, Mfr. By: Bradshaw Medical... on . The reason FDA records is "The firm learned that the Torque Driver Handles were not actuating at a proper torque. It was determined that the handles were speed determinant and could output a higher torque if the handles were turned at a very slow...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0034-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Xtant Medical Holdings, Inc.
Product
Axle Interspinous Fusion System Torque Driver Handle, Model #: X060-0323, Lot # 525941, S/Ns 001-017, Lot # 525942, S/Ns 001-050, Mfr. By: Bradshaw Medical Inc., Kenosha, WI 53144 The Axle Torque Driver Handle is used to lock the set screw into the locking plate.
Product code
FZS Curette, Surgical, General Use
Reason for recall
The firm learned that the Torque Driver Handles were not actuating at a proper torque. It was determined that the handles were speed determinant and could output a higher torque if the handles were turned at a very slow speed.
Root cause tag
design
FDA root cause
Component design/selection
Action
X-spine Systems, Inc. sent an URGENT DEVICE RECALL letter dated September 12, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to examine their stock immediately to determine if they have any affected product. If so, customers were instructed once the set screw is tightened onto the locking plate, and resistance is detected in the handle, it will take less than 2 seconds to hear the handle "click." The "click" indicates the torque limit was reached. If further resistance is felt after 2 seconds, the surgeon should release the handle from tension, and try again. If the product was distributed further, they should contact their accounts, and advise them of this recall. The firm will send replacement handles. For any questions call (937) 847-8400, ext 115.
Quantity in commerce
39 Torque Driver Handles
Distribution
Nationwide Distribution including CA, CO, FL, IA, IN, LA, NY, PA, TN, TX and UT.
Product codes and lots
Model #: X060-0323, Lot # 525941, S/Ns 001-017, Lot # 525942, S/Ns 001-050