Z-3076-2011
Hamamatsu model L10951-35 Micro Focus x-ray generator. The x-ray generator is incorporated into a cabinet...
Hamamatsu Corp. / initiated 2011-07-07 / Terminated
Hamamatsu Corporation began a recall of Hamamatsu model L10951-35 Micro Focus x-ray generator. The x-ray generator is incorporated into a cabinet x-ray system that is used for a variety of studies on . The reason FDA records is "It was discovered about a potential issue of X-ray leakage with our Model # LI1 0951-35 Micro Focus X-Ray generator". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-3076-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Hamamatsu Corp.
- Product
- Hamamatsu model L10951-35 Micro Focus x-ray generator. The x-ray generator is incorporated into a cabinet x-ray system that is used for a variety of studies.
- Product code
- RBS Industrial X-Ray
- Reason for recall
- It was discovered about a potential issue of X-ray leakage with our Model # LI1 0951-35 Micro Focus X-Ray generator.
- FDA root cause
- Radiation Control for Health and Safety Act
- Action
- Hamamatsu Corporation sent a letter dated July 19, 2011, to the manufacturer and their subcontractor informing them of the affected product and what they propose to do. They will repair the units both in the field and those returned to our plant by installing additional lead shielding in the units. The lead shielding will be fixed to the unit by soldering the addition to the existing lead shield (see attachment 2). We expect to complete the repair of all units in the field by October 1, 2011. All new units to be shipped from our plant will be repaired before shipment. For further questions, please call (908) 231-0960 X 2300
- Quantity in commerce
- 8
- Distribution
- Worldwide Distribution -- USA ( nationwide ) the countries of Aichi, India and Japan.
- Product codes and lots
- Model # L10951-35 Serial number SII00002 through SII00009.