Z-3076-2011

Hamamatsu model L10951-35 Micro Focus x-ray generator. The x-ray generator is incorporated into a cabinet...

Hamamatsu Corp. / initiated 2011-07-07 / Terminated

Hamamatsu Corporation began a recall of Hamamatsu model L10951-35 Micro Focus x-ray generator. The x-ray generator is incorporated into a cabinet x-ray system that is used for a variety of studies on . The reason FDA records is "It was discovered about a potential issue of X-ray leakage with our Model # LI1 0951-35 Micro Focus X-Ray generator". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-3076-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Hamamatsu Corp.
Product
Hamamatsu model L10951-35 Micro Focus x-ray generator. The x-ray generator is incorporated into a cabinet x-ray system that is used for a variety of studies.
Product code
RBS Industrial X-Ray
Reason for recall
It was discovered about a potential issue of X-ray leakage with our Model # LI1 0951-35 Micro Focus X-Ray generator.
FDA root cause
Radiation Control for Health and Safety Act
Action
Hamamatsu Corporation sent a letter dated July 19, 2011, to the manufacturer and their subcontractor informing them of the affected product and what they propose to do. They will repair the units both in the field and those returned to our plant by installing additional lead shielding in the units. The lead shielding will be fixed to the unit by soldering the addition to the existing lead shield (see attachment 2). We expect to complete the repair of all units in the field by October 1, 2011. All new units to be shipped from our plant will be repaired before shipment. For further questions, please call (908) 231-0960 X 2300
Quantity in commerce
8
Distribution
Worldwide Distribution -- USA ( nationwide ) the countries of Aichi, India and Japan.
Product codes and lots
Model # L10951-35 Serial number SII00002 through SII00009.