Z-0338-2012
Incu i (Atom Infant Incubator Model 101) Atom Medical Corporation 3-18-15 Hongo, Bunkyo-Ku, Tokyo, Japan...
Atom Medical Corporation / initiated 2011-08-23 / Terminated / root cause: design
Atom Medical Corporation began a recall of Incu i (Atom Infant Incubator Model 101) Atom Medical Corporation 3-18-15 Hongo, Bunkyo-Ku, Tokyo, Japan 113-0033 Usage: An incubator for infants on . The reason FDA records is "The caster with lock on the incubator may become loose or break". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0338-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Atom Medical Corporation
- Product
- Incu i (Atom Infant Incubator Model 101) Atom Medical Corporation 3-18-15 Hongo, Bunkyo-Ku, Tokyo, Japan 113-0033 Usage: An incubator for infants.
- Product code
- FMZ Incubator, Neonatal
- Reason for recall
- The caster with lock on the incubator may become loose or break.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- The firm, Atom Medical, sent an "Urgent: Field Safety Notice" letter dated August 18, 2011, to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to immediately identify all devices in their inventory and quarantine them until the rework is complete; identify any customers to whom they have further distributed the product; contact their customers at once and arrange a time in which the field service technician can perform the rework (the parts needed to complete the rework will be sent from Atom Medical's export department ); complete rework of all affected devices and document rework on the Distributor FSCA/Recall Response Form and return via Fax to: +81 3 3812-3199 or email to [email protected]; [email protected] or [email protected] (all forms should be completed and returned by January 31, 2012.) If you have any questions, contact the Manager Quality Control or Manager, Regulatory Affairs at email addresses above, phone: +81-3-3815-3632 or fax +81-3-3812-3199.
- Distribution
- Worldwide distribution: USA (nationwide) including state of: PA and countries of: Algeria, Austria, Australia, China, Egypt, Germany, Italy, Japan, Korea, Kuwait, Malaysia, Portugal, Russia, Saudi Arabia, Spain, Switzerland, Thailand, The Netherlands, and UAE.
- Product codes and lots
- All lots on the market. model 101