BD Microtainer MAP Microtube for Automated Process with K2EDTA. For the collection, transport, and automated...
Becton Dickinson & Company began a recall of BD Microtainer MAP Microtube for Automated Process with K2EDTA. For the collection, transport, and automated processing of skin punctures samples from infants... on . The reason FDA records is "BD Microtainer MAP Microtube is being recalled due to the potential clotting of blood specimens resulting from a reduced amount of EDTA additive". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0176-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Becton Dickinson & Co
- Product
- BD Microtainer MAP Microtube for Automated Process with K2EDTA. For the collection, transport, and automated processing of skin punctures samples from infants, children, geriatric, oncology, and critical care patients.
- Product code
- JKA Tubes, Vials, Systems, Serum Separators, Blood Collection
- Reason for recall
- BD Microtainer MAP Microtube is being recalled due to the potential clotting of blood specimens resulting from a reduced amount of EDTA additive.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- A Becton Dickinson Sales Consultant conducted a "on-site" visit to the single Customer account, Florida Hospital, on February 11, 2011. All units were successfully recovered. All product recovered was quarantined and destroyed. Becton Dickinson believes that all reasonable efforts to communicate the issue to the affected consignee have been completed. If you have any questions or request additional information please do not hesitate to call (201) 847-5636.
- Quantity in commerce
- 10,800 units
- Distribution
- Distributed only to Florida.
- Product codes and lots
- Lot # 1018904 Catalog # 363706