Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-3274-2011

Stryker SmartPump Extension Tubing (Pneumatic tourniquet). Sterile tubing extension 40" and 60", set, luer...

Stryker Corp. / initiated 2011-08-25 / Terminated / root cause: labeling and UDI

Stryker Instruments Div. of Stryker Corporation began a recall of Stryker SmartPump Extension Tubing (Pneumatic tourniquet). Sterile tubing extension 40" and 60", set, luer lock. Rx only. Stryker Instruments Div. of Stryker... on . The reason FDA records is "Three complaints have been reported indicating that the product inside the package did not match the label description. The units had quick connectors instead of luer lock connectors as stated in the label description". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-3274-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Stryker Corp.
Product
Stryker SmartPump Extension Tubing (Pneumatic tourniquet). Sterile tubing extension 40" and 60", set, luer lock. Rx only. Stryker Instruments Div. of Stryker Corporation. Optional accessories used with the Stryker SmartPump Tourniquet System. The luer locks are intended to be connected to the tourniquet pump - to fill the line - and to the cuff. The extension tubing is an optional accessory used to connect the tourniquet cuff with the pump when a longer distance between the pump and the tourniquet is desired. The extension tubing comes packaged 10 per box.
Product code
KCY Tourniquet, Pneumatic
Reason for recall
Three complaints have been reported indicating that the product inside the package did not match the label description. The units had quick connectors instead of luer lock connectors as stated in the label description.
Root cause tag
labeling and UDI
FDA root cause
Mixed-up of materials/components
Action
Stryker Instruments sent out "Urgent Medical Device Recall" notices dated August 22, 2011 to all affected customers. The letter included the name of the recalled product, lot numbers, and the reason for the recall. Customers were instructed to send back any product to Stryker that is within the affected lots. The letter further stated Stryker will quarantine the product and replace any product that is returned. For information on this recall contact Stryker Instruments at (269) 389-4534.
Quantity in commerce
5920-000-003 (30 bx) and 5920-000-005 (30 bx)
Distribution
Worldwide Distribution: Nationwide Distribution including the states of HAWAII, NORTH DAKOTA, NEBRASKA, PENNSYLVANIA, and IOWA, and the country of FRANCE
Product codes and lots
60" Part number 5920-000-003; lot number 11137012. 40" Part number 5920-000-005; lot number 11138012.