Straumann coDiagnostiX, version 8.0, used to guide a clinician in the proper implant selection, oral...
Straumann Manufacturing Inc. began a recall of Straumann coDiagnostiX, version 8.0, used to guide a clinician in the proper implant selection, oral placement and the proper selection of an oral drill... on . The reason FDA records is "Software version 8.0 has a software anomaly, in which the length is not adjusted accordingly when it is changed after the surgical sleeve position planning step has been completed". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-3184-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Straumann USA, LLC
- Product
- Straumann coDiagnostiX, version 8.0, used to guide a clinician in the proper implant selection, oral placement and the proper selection of an oral drill. Imaging processing System
- Product code
- LLZ System, Image Processing, Radiological
- Reason for recall
- Software version 8.0 has a software anomaly, in which the length is not adjusted accordingly when it is changed after the surgical sleeve position planning step has been completed.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Staumann USA notified accounts by an Urgent Field Safety Notice letter dated August 4, 2011 and delivered via Federal Express Delivery Service. The letter identified the affected product and explained the reason for the recall. Customers were asked to review the letter, take appropriate actions, including completing and faxing the enclosed. Customer Confirmation Form. Straumann released a corrected software upgrade (coDiagnostiX 8.0.2.3672), which is available for download when users go online and activate current software. If the user does not have internet access a replacement CD can be ordered from Technical Support at 866 - 531 -7365. Questions should be directed to Technical Support Monday through Friday from 8:00 am to 8:00 pm Eastern Time.
- Quantity in commerce
- 272
- Distribution
- Nationwide Distribution
- Product codes and lots
- Software version 8.0 as listed below: listed are the affected articles. Internet Upgrade: 019,0385 @ Upgrade cDX 8 PPP (7+) 019,0386 @ Upgrade cDX 8 Station (7+) 019.0387 @ Upgrade cDx 8 coDoctor (7+) 019.0388 @ Upgrade cDX coDoctor (7+) > PPP 8 0'19.0389 @ Upgrade cDX codoctor (7+) > PPG 8 019.0390 @ Upgrade cDX coDoctor (7+) >Station 8 019.0391 @ Upgrade cDX PPP (7 +):> PPG 8 019.0392 @ Upgrade cDX PPP (7+):> Station 8 019.0393 @ Upgrade coDX PPP > Station 8 Offline Upgrades: 019.0378 Upgrade coDiagnostix 8 PPP (7-) 0'19.0379 Upgrade coDiagnostiX 8 Station (7-) 019.0380 Upgrade coDiagnostiX 8 coDoctor (7-) 019.0381 Upgrade coDoctor (7-) -> PPP 8 019.0382 Upgrade coDoctor (7-) -> PPG8 019.0383 Upgrade PPP (7-) -> PPG 8 019.0384 coDiagnostiX CD coDlagnostiX licenses: 019.0370 coDiagnostiX 8 PPP 019.0371 coDiagnostiX 8 PPG 0'19.0372 coDiagnostiX 8 Station, and 019.0373 coDiagnositX 8 coDoctor.