Z-0327-2012

GammaMed Plastic Needle with Mandrin; 2.0 mm diameter, length 200 mm. Remote controlled radionuclide...

Varian Medical Systems, Inc. / initiated 2008-09-30 / Terminated / root cause: manufacturing process

Varian Medical Systems, Inc. began a recall of GammaMed Plastic Needle with Mandrin; 2.0 mm diameter, length 200 mm. Remote controlled radionuclide applicator system. Varian Medical Systems, Inc. Mfg by... on . The reason FDA records is "Plastic needle tip could possibly crack or separate". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0327-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Varian Medical Systems, Inc.
Product
GammaMed Plastic Needle with Mandrin; 2.0 mm diameter, length 200 mm. Remote controlled radionuclide applicator system. Varian Medical Systems, Inc. Mfg by Varian Medical Systems Haan GmbH, Haan, Germany. The interstitial plastic needles with 2 mm diameter are designed for interstitial treatments in areas such as head and neck, gynecology, breast or prostate. In order to avoid bending and to facilitate the introduction of the needles, each needle is supplied with a mandrin.
Product code
JAQ System, Applicator, Radionuclide, Remote-Controlled
Reason for recall
Plastic needle tip could possibly crack or separate.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Varian Medical Systems sent a "CUSTOMER TECHNICAL BULLETIN" letter dated September 30, 2008 to all affected customers. The letter described the product, problem, and actions to be taken by the customers. The letter instructed users to inspect needle tips before and after every clinical application to prevent needle tip separation or cracking.
Quantity in commerce
1716 parts
Distribution
Worldwide Distribution - USA (nationwide) including the states of Arizonia, California, Florida, Illinois, Indiana, Michigan, North Carolina, and Vermont, and the countries of Argentina, Austria, Belgium, Canada, Estonia, France, Germany, India, Norway, Pakistan, Russia, Slovenia, So. Africia, Spain, Switzerland, United Kingdom, and Ukraine.
Product codes and lots
Part number GM11007570