Z-0328-2012
GammaMed Plastic Needle with Mandrin; 2.0 mm diameter, length 320 mm. Remote controlled radionuclide...
Varian Medical Systems, Inc. / initiated 2008-09-30 / Terminated / root cause: manufacturing process
Varian Medical Systems, Inc. began a recall of GammaMed Plastic Needle with Mandrin; 2.0 mm diameter, length 320 mm. Remote controlled radionuclide applicator system. Varian Medical Systems, Inc. Mfg by... on . The reason FDA records is "Plastic needle tip could possibly crack or separate". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0328-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Varian Medical Systems, Inc.
- Product
- GammaMed Plastic Needle with Mandrin; 2.0 mm diameter, length 320 mm. Remote controlled radionuclide applicator system. Varian Medical Systems, Inc. Mfg by Varian Medical Systems Haan GmbH, Haan, Germany. The interstitial plastic needles with 2 mm diameter are designed for interstitial treatments in areas such as head and neck, gynecology, breast or prostate. In order to avoid bending and to facilitate the introduction of the needles, each needle is supplied with a mandrin.
- Product code
- JAQ System, Applicator, Radionuclide, Remote-Controlled
- Reason for recall
- Plastic needle tip could possibly crack or separate.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Varian Medical Systems sent a "CUSTOMER TECHNICAL BULLETIN" letter dated September 30, 2008 to all affected customers. The letter described the product, problem, and actions to be taken by the customers. The letter instructed users to inspect needle tips before and after every clinical application to prevent needle tip separation or cracking.
- Quantity in commerce
- 1716 parts
- Distribution
- Worldwide Distribution - USA (nationwide) including the states of Arizonia, California, Florida, Illinois, Indiana, Michigan, North Carolina, and Vermont, and the countries of Argentina, Austria, Belgium, Canada, Estonia, France, Germany, India, Norway, Pakistan, Russia, Slovenia, So. Africia, Spain, Switzerland, United Kingdom, and Ukraine.
- Product codes and lots
- Part number GM11007580