Z-0015-2012

Custom anesthesia kits labeled as follows: (1) DeRoyal (R) Adult Exp Anesthesia Circuit, REF 86-001737, Rx...

Deroyal Industries, Inc. Lafollette / initiated 2011-08-23 / Terminated / root cause: manufacturing process

Deroyal Industries, Inc. Lafollette began a recall of Custom anesthesia kits labeled as follows: (1) DeRoyal (R) Adult Exp Anesthesia Circuit, REF 86-001737, Rx Only, NON-STERILE, Manufactured by: DeRoyal... on . The reason FDA records is "The firm manufactured kits containing a HCH device, manufactured by Vital Signs, that was subsequently recalled because the product may have an occlusion that could prevent proper flow of medical gases or oxygen to the...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0015-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Deroyal Industries, Inc. Lafollette
Product
Custom anesthesia kits labeled as follows: (1) DeRoyal (R) Adult Exp Anesthesia Circuit, REF 86-001737, Rx Only, NON-STERILE, Manufactured by: DeRoyal Industries, Inc. 200 DeBusk Lane, Powell, TN 37849, USA; Lot Number 25660996 (2) DeRoyal (R) Adult Anesthesia Circuit, REF 86-001781, Rx Only, NON-STERILE, Manufactured by: DeRoyal Industries, Inc. 200 DeBusk Lane, Powell, TN 37849, USA; Lot Numbers: 25574115, 25624514, 25661260, 25702578, 25766272, 25835901, 25835910, 25873406, 25906061, 25960681 (3) DeRoyal (R) Latex Safe Anesthesia Kit, REF 86-001641, Rx Only, NON-STERILE, Manufacturer: DeRoyal Industries, Inc. 200 DeBusk Lane, Powell, TN 37849, USA: Lot Numbers: 25573171, 25765667, 25834853, 25834861 (4) DeRoyal (R) Custom Anesthesia Kit, REF 86-001619, Rx Only, NON-STERILE, Manufacturer: DeRoyal Industries, Inc. 200 DeBusk Lane, Powell, TN 37849, USA: Lot Number 25573155 (5) DeRoyal (R) Anesthesia Kit, REF 86-001669, Rx Only, NON-STERILE, Manufacturer: DeRoyal Industries, Inc. 200 DeBusk Lane, Powell, TN 37849, USA: Lot Numbers: 25313484, 25702324, 25977977 Intended Use: Varies by kit; an anesthesia breathing circuit is a device that is intended to administer medical gases to a patient during anesthesia. It provides both an inhalation and exhalation route and may include a connector, adaptor, and Y-piece.
Product code
OGD Custom Anesthesia Tray
Reason for recall
The firm manufactured kits containing a HCH device, manufactured by Vital Signs, that was subsequently recalled because the product may have an occlusion that could prevent proper flow of medical gases or oxygen to the patient.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
The firm, DeRoyal, sent a "URGENT: MEDICAL DEVICE RECALL" letter dated September 13, 2011 to its customers. The letter describes product, problem, and actions to be taken. Customers are instructed to identify inventory being recalled and place in quarantine to prevent further use. Constomers were asked to return affected product they have in inventory. DeRoyal will replace this product free of charge. Customers were asked to complete the attached Recall Response Spreadsheet and return as per the instructions. Constomers were asked to return affected product they have in inventory. DeRoyal will replace this product free of charge. Distributors were instructed to perform a sub-recall to their customers. Customers with questions should call 865-362-1037 between 9am and 5pm EST or email [email protected].
Quantity in commerce
940 units
Distribution
Nationawide Distribution USA - including the states of: CA, FL, GA, NJ, and WA.
Product codes and lots
REF 86-001737, Lot Number 25660996 REF 86-001781, Lot Numbers: 25574115, 25624514, 25661260, 25702578, 25766272, 25835901, 25835910, 25873406, 25906061, 25960681 REF 86-001641, Lot Numbers: 25573171, 25765667, 25834853, 25834861 REF 86-001619, Lot Number 25573155 REF 86-001669, Lot Numbers: 25313484, 25702324, 25977977