Z-3271-2011

VASOVIEW 7xB Endoscopic Vessel Harvesting System; Manufactured by Maquet Cardiovascular, LLC; 170 Baytech...

Maquet Cardiovascular LLC / initiated 2009-12-17 / Terminated / root cause: manufacturing process

Maquet Cardiovascular, LLC began a recall of VASOVIEW 7xB Endoscopic Vessel Harvesting System; Manufactured by Maquet Cardiovascular, LLC; 170 Baytech Drive, San Jose, CA 95134. Product Usage: Endoscopic... on . The reason FDA records is "During routine production quality monitoring process, it was found that UV curing adhesive was not used in a limited number of devices". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-3271-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Maquet Cardiovascular LLC
Product
VASOVIEW 7xB Endoscopic Vessel Harvesting System; Manufactured by Maquet Cardiovascular, LLC; 170 Baytech Drive, San Jose, CA 95134. Product Usage: Endoscopic vessel harvesting systems are indicated for use in minimally invasive surgery allowing access for vessel harvesting and are primarily indicated for patients undergoing endoscopic surgery for arterial bypass.
Product code
GEI Electrosurgical, Cutting & Coagulation & Accessories
Reason for recall
During routine production quality monitoring process, it was found that UV curing adhesive was not used in a limited number of devices.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
MAQUET Cardiovascular, LLC, sent an "Urgent Device Removal/Immediate Action Required" letter dated December 17, 2009, to all affected customers. The letter identified the recalled products, the problem, and the actions needed to be taken. The letters asked customers to examine their stocks (inventory) immediately to determine if they had any affected lot numbers and to discontinue dispensing (distributing) the lot. The letter states that customers are to complete the attached Field Action Response form regardless of whether or not the affected products are located in their inventory. Distributors are instructed to provide the same instructions to their customers supplied with the recalled products. The letter instructed all users to return all unused affected products to MAQUET. For customer's convenience, all affected products that are returned may be exchanged with other unaffected VASOVIEW Endoscopic Vessel Harvesting System by overnight delivery. Questions should be directed to MAQUET Cardiovascular sales representative or Customer Service at 1-888-880-2874.
Quantity in commerce
177 units
Distribution
Nationwide Distribution (USA) including the states of: AL, AZ, CA, CT, FL, KS, KY, IL, IN, MI, MO, ND, NY, NV, OH, OK, OR, PA, SC, TN, TX, UT, WA, and WI.
Product codes and lots
Part number: VH-3200; Lot # 9092371, Exp 8/31/2010; Lot # 9092571, Exp 8/31/2010 - never shipped.