Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0136-2012

SwitchPoint Infinity Control System Stryker Communications, 1410 Lakeside Parkway, # 100, Flower Mound, TX...

STRYKER Communications, Inc. / initiated 2011-08-23 / Terminated / root cause: manufacturing process

Stryker Communications Corp began a recall of SwitchPoint Infinity Control System Stryker Communications, 1410 Lakeside Parkway, # 100, Flower Mound, TX 75208 The SwitchPoint Infinity (SPI) Control System... on . The reason FDA records is "Installation records for the SwitchPoint Infinity, Booms, Lights, and Flat Panel systems are incomplete or inadequate". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0136-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
STRYKER Communications, Inc.
Product
SwitchPoint Infinity Control System Stryker Communications, 1410 Lakeside Parkway, # 100, Flower Mound, TX 75208 The SwitchPoint Infinity (SPI) Control System is an integrated voice, video, data router and teleconferencing interface for operating rooms. The SwitchPoint Infinity is used for direct control of room equipment, audio/video equipment, and indirect control of surgical equipment in the operating room.
Product code
FSY Light, Surgical, Ceiling Mounted
Reason for recall
Installation records for the SwitchPoint Infinity, Booms, Lights, and Flat Panel systems are incomplete or inadequate.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Stryker sent an URGENT FIELD SAFETY NOTICE: RA2010-245 dated August 23, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to complete the attached Customer Response Form within 5 business days to Stryker. A Stryker representative would contact them at their convenience to schedule a re-inspection for all of the units identified and still in use. In the event the inspected equipment does not meet Stryker's established specifications, their representative will visit their facility and hopefully resolve the issue. If a subsequent visit is required the date will be scheduled in full consideration of their services. For questions regarding this recall call 972-410-7310.
Distribution
Worldwide Distribution including Puerto Rico, Brazil, Ecuador, Venezuela, Costa Rico, Bolivia, Peru, Mexico, China, India, Korea, Philippines, Singapore, Malaysia, Taiwan, Middle East, Russia, Africa, Poland, South Africa, Australia, New Zealand, Canada, Japan, Denmark, Finland, Norway, Sweden, Belgium, The Netherlands, France, Germany, Austria, Italy, Portugal, Spain, Romania, Switzerland, and United Kingdom.
Product codes and lots
Catalog numbers: 0678000530, 0678000531, 0678000532, 0678000533, 0678000534, 0678000535, 0678000536, and 0678000537. Serial Numbers: All product distributed and installed between July 2004 and March 2011.