Z-0020-2012

VARIANT II TURBO Hemoglobin A1c Program Reorder Pack, 1600 tests, IVD, Made in United States. Kit contains...

Bio-Rad Laboratories Inc. / initiated 2011-09-12 / Terminated / root cause: manufacturing process

Bio-Rad Laboratories, Inc. began a recall of VARIANT II TURBO Hemoglobin A1c Program Reorder Pack, 1600 tests, IVD, Made in United States. Kit contains: 1. Elution Buffer A 3 x 2500 mL, Bis-Tris Phosphate... on . The reason FDA records is "A previous recall related to reports of calibration failure and late retention times using Elution Buffer led to an examination of other buffers manufactured during the same time period and a decision was made to field...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0020-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Bio-Rad Laboratories Inc.
Product
VARIANT II TURBO Hemoglobin A1c Program Reorder Pack, 1600 tests, IVD, Made in United States. Kit contains: 1. Elution Buffer A 3 x 2500 mL, Bis-Tris Phosphate Buffer 2. Elution Buffer B 1 x 1500 mL Bis-Tris Phosphate Buffer 3. Cartridge Set, 2 cation exchange analytical cartridges, 4 cation exchange Guard cartridges. 4. CD-ROM 1 - CD Rom with program parameters 5. Calibration/Diluent set - 2 levels, 2 vial each of lyophilized human red blood cell hemolysate with preservative, 1 bottle diluent, 100 mL deionized water. 6. Whole Blood Primer, 6 x 1 mL, lyophilized human red blood cell hemolysate with preservative, 7. Sample vials - 2 x 100, 1.5 mL polypropylene vials with pierceable caps. 8. Instruction Manual. Bio-Rad Laboratories, Hercules, CA 94547. Percentage determination of hemoglobin A1c in human whole blood using ion-exchange high=performance liquid chromatography (HPLC)
Product code
LCP Assay, Glycosylated Hemoglobin
Reason for recall
A previous recall related to reports of calibration failure and late retention times using Elution Buffer led to an examination of other buffers manufactured during the same time period and a decision was made to field correct these products until a definitive root cause and corrective action have been found and implemented.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Bio-Rad sent an Urgent Medical Device Correction letter on September 12, 2011, to all affected subsidiaries and customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to discard all affected product in accordance with all local, state and federal disposal procedures. If customers distributed the product further they should direct their customers to discard all unused product. Customers were asked to please translate (if necessary) the attached "Customer Medical Device Correction Response Form," customers should add their local contact information and distribute to all affected customers in their region by mail or fax. Compile tracking documentation (e.g. Excel spreadsheet) for all customers who received the Medical Device Correction Letter. Complete the "SUBSIDIARY MEDICAL DEVICE CORRECTION RESPONSE FORM" and fax to Bio-Rad Regulatory Affairs at (510) 741-3954, or e-mail PDF to CSD_RA/Hercules/US/BIO-RAD. Customers would be issued credits for the affected products based on their request. For questions call (510) 741-3954.
Quantity in commerce
82
Distribution
Nationwide and to New Zealand, Canada, France, England, China, India, Mexico, Italy, Brazil, Australia, Singapore, Thailand, SriLanka, hong Kong, Sweden, Germany, Peru, Jamaica, Uruguay, Columbia, Argentina, Chile, Paraguay and Korea.
Product codes and lots
Model number: 270-2417; Lot number 70211073, expiry 12/31/2011. Contains Buffer lot: AA10758/AA10759, expiry 12/31/11.