Custom Dialysis prep kits labeled as follows: ADV-CARE-ISL44, 100/CS, LOT #K1018, EXP 5-12, BIN 34B...
Lee Medical International Inc began a recall of Custom Dialysis prep kits labeled as follows: ADV-CARE-ISL44, 100/CS, LOT #K1018, EXP 5-12, BIN 34B; COMMUNITY-CATH, 150/CS, LOT #K1052, EXP 5-12, BIN 36B... on . The reason FDA records is "The firm used alcohol triple swabstick and prep pads manufactured by H & P Industries in the manufacture of their custom dialysis kits. Those swabsticks and pads were subsequently recalled due to possible bacterial...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-3260-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Lee Medical International Inc
- Product
- Custom Dialysis prep kits labeled as follows: ADV-CARE-ISL44, 100/CS, LOT #K1018, EXP 5-12, BIN 34B; COMMUNITY-CATH, 150/CS, LOT #K1052, EXP 5-12, BIN 36B; COLUMBUS-CATH, 50/CS, LOT #1080, EXP 05-12, BIN 39B; RENALSG-CATH, 75/CS, LOT #L1046, EXP 6-12, BIN 40B; SUNYDOWN-HEMO, 100/CS, LOT #l1015, EXP 6-12, BIN 28A; KIDNEYCARE 125/CS, LOT #l1070, EXP 6-12; METHODIST-CATH, 75/CS, LOT #L1072, EXP 6-12; METHODIST-HEMO, 150/CS, LOT #L1073, EXP 6-12; MERCYMED-CATH, 75/CS, LOT #L1077, EXP 6-12, BIN 39A and SANTACLA-CATH, 75/CS, LOT #A1118, EXP 7-12, BIN 44A. Product Usage: custom dialysis prep kits
- Product code
- FKG Tray, Start/Stop (Including Contents), Dialysis
- Reason for recall
- The firm used alcohol triple swabstick and prep pads manufactured by H & P Industries in the manufacture of their custom dialysis kits. Those swabsticks and pads were subsequently recalled due to possible bacterial contamination.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- The firm, Lee Medical International Inc., notified their consignees by email and telephone on August 17, 2011. Consignees were instructed to remove the affected product from stock, quarantine for re-processing, and return or destroy product. Lee Medical arranged for replacement product from alternate manufacturer to be sent to customers. For questions call Production Manager or Operations Manager at 504-734-9336 or email [email protected] or [email protected].
- Quantity in commerce
- 4,075 units
- Distribution
- Nationwide Distribution (USA) including the states of: CA, FL, GA, IA, LA, MD, NY and TX.
- Product codes and lots
- ADV-CARE-ISL44, 100/CS, LOT #K1018, EXP 5-12, BIN 34B; COMMUNITY-CATH, 150/CS, LOT #K1052, EXP 5-12, BIN 36B; COLUMBUS-CATH, 50/CS, LOT #1080, EXP 05-12, BIN 39B; RENALSG-CATH, 75/CS, LOT #L1046, EXP 6-12, BIN 40B; SUNYDOWN-HEMO, 100/CS, LOT #l1015, EXP 6-12, BIN 28A; KIDNEYCARE 125/CS, LOT #l1070, EXP 6-12; METHODIST-CATH, 75/CS, LOT #L1072, EXP 6-12; METHODIST-HEMO, 150/CS, LOT #L1073, EXP 6-12; MERCYMED-CATH, 75/CS, LOT #L1077, EXP 6-12, BIN 39A and SANTACLA-CATH, 75/CS, LOT #A1118, EXP 7-12, BIN 44A.