Sodium Bicarbonate Mixing and Delivery System Software v 7.1. Model # MD-380, Serial Numbers: 5103031102...
Isopure Corp began a recall of Sodium Bicarbonate Mixing and Delivery System Software v 7.1. Model # MD-380, Serial Numbers: 5103031102, 5101101101, 5102161007, 5103221007, 5104221007... on . The reason FDA records is "Software problem causes transfer line between mixing and storage tank to not be rinsed after disinfection, leaving residual chlorine in the line". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0033-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Isopure Corp
- Product
- Sodium Bicarbonate Mixing and Delivery System Software v 7.1. Model # MD-380, Serial Numbers: 5103031102, 5101101101, 5102161007, 5103221007, 5104221007, 5105101008, 5109131010, 5111231011, 5114301012, 5102131101 & 5113091012. Manufactured by: Isopure Corp, Simpsonville, KY System is used to mix and distribute bicarbonate as a part of a clinical dialysis system.
- Product code
- FJP Adaptor, Y
- Reason for recall
- Software problem causes transfer line between mixing and storage tank to not be rinsed after disinfection, leaving residual chlorine in the line.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Isopure sent a letter dated May 19, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Isopure devised an immediate solution which must be implemented at their faclity in conjunction with the disinfection procedure. Customers were asked to inform all appropriate personnel, clinical and technical, who may be involved in operating the system during the normal disinfection schedule. On 5/19/2011, the firm sent letters to their customers via US mail and email. They also telephoned their customers to notify them of the problem.
- Quantity in commerce
- 11 units
- Distribution
- Nationwide Distribution including AR, CT, GA, MA, ME, NJ, PA, TN and WV.
- Product codes and lots
- Model # MD-380, Serial Numbers: 5103031102, 5101101101, 5102161007, 5103221007, 5104221007, 5105101008, 5109131010, 5111231011, 5114301012, 5102131101 & 5113091012.