Z-0325-2012

VariSource iX, Software Version 1.1, Part Number AL20020000 VariSource iX Afterloader. The VariSource iX...

Varian Medical Systems, Inc. / initiated 2010-03-01 / Terminated / root cause: software

Varian Medical Systems, Inc. began a recall of VariSource iX, Software Version 1.1, Part Number AL20020000 VariSource iX Afterloader. The VariSource iX Series High Dose Rate Afterloader system is a computer... on . The reason FDA records is "VariSource iX console software may cause incorrect dwell times to be delivered to patient and may result in a mistreatment". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0325-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Varian Medical Systems, Inc.
Product
VariSource iX, Software Version 1.1, Part Number AL20020000 VariSource iX Afterloader. The VariSource iX Series High Dose Rate Afterloader system is a computer controlled remote electra/mechanical system used for medical purposes, for placing a NiTiNol wire incorporating an irradiated iridium seed internally or close by, a malignant tumor or tumor bed in a practice known as brachytherapy. The device has up to 20 channels for treatment.
Product code
JAQ System, Applicator, Radionuclide, Remote-Controlled
Reason for recall
VariSource iX console software may cause incorrect dwell times to be delivered to patient and may result in a mistreatment.
Root cause tag
software
FDA root cause
Software design
Action
Varian Medical Systems sent a "MEDICAL DEVICE RECALL" letter dated April 1, 2010 to all affected customers. The letter describes the product, problem, and actions to be taken by the customers. Customers were asked to conduct a simple test to establish if their machine is affected by the problem. Device users were requested to complete an attached receipt verification form and return it to Varian BrachyTherapy via fax or e-mail. Call the Contact Service Dispatch number at 1-800-864-1672 for questions regarding this detail.
Quantity in commerce
72 worldwide/50 in U.S.
Distribution
Worldwide Distribution-USA (nationwide)
Product codes and lots
Serial numbers: 212, 215, 222, 224, 239, 246, 271, 274, 281, 293, 308, 322, 333, 343, 346, 388, 394, 401, 403, 406, 411, 431, 436, 437, 438, 441, 444, 446, 450, 456, 462, 500, 501, 505, 509, 510, 511, 512, 513, 514, 515, 516, 518, 519, 521, 522, 527, 530, 532, 535, 536, 538, 539, 540, 541, 542, 544, 545, 546, 548, 551. Part Number: AL20020000 (VariSource iX Afterloader