Z-0069-2012
Siemens Healthcare Diagnostics, Inc. Dimension(R) EXL Cardiac Troponin I Flex(R) Reagent Cartridges
Siemens Healthcare Diagnostics Inc. / initiated 2011-08-31 / Terminated / root cause: manufacturing process
Siemens Healthcare Diagnostics, Inc. began a recall of Siemens Healthcare Diagnostics, Inc. Dimension(R) EXL Cardiac Troponin I Flex(R) Reagent Cartridges on . The reason FDA records is "Firm has confirmed customer complaints for low recovery of QC and patient samples". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0069-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Siemens Healthcare Diagnostics, Inc. Dimension(R) EXL Cardiac Troponin I Flex(R) Reagent Cartridges
- Product code
- JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Reason for recall
- Firm has confirmed customer complaints for low recovery of QC and patient samples.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Siemens Healthcare Diagnostics, Inc. Urgent Field Safety Notice was sent to customers on August 31, 2011 regarding the issue. The letter instructs customers to immediately discontinue use of the affected lots and discard all remaining inventory.
- Quantity in commerce
- 6,072 cartons
- Distribution
- Worldwide distribution: USA, Puerto Rico, Guam, Austria, Canada, Croatia, France, Greece, Germany, Italy, Japan, Mexico, Portugal, Spain and Switzerland.
- Product codes and lots
- Lot Numbers ED2098, EA2111, FB2134, ED2147