Z-0069-2012

Siemens Healthcare Diagnostics, Inc. Dimension(R) EXL Cardiac Troponin I Flex(R) Reagent Cartridges

Siemens Healthcare Diagnostics Inc. / initiated 2011-08-31 / Terminated / root cause: manufacturing process

Siemens Healthcare Diagnostics, Inc. began a recall of Siemens Healthcare Diagnostics, Inc. Dimension(R) EXL Cardiac Troponin I Flex(R) Reagent Cartridges on . The reason FDA records is "Firm has confirmed customer complaints for low recovery of QC and patient samples". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0069-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Siemens Healthcare Diagnostics, Inc. Dimension(R) EXL Cardiac Troponin I Flex(R) Reagent Cartridges
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
Firm has confirmed customer complaints for low recovery of QC and patient samples.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Siemens Healthcare Diagnostics, Inc. Urgent Field Safety Notice was sent to customers on August 31, 2011 regarding the issue. The letter instructs customers to immediately discontinue use of the affected lots and discard all remaining inventory.
Quantity in commerce
6,072 cartons
Distribution
Worldwide distribution: USA, Puerto Rico, Guam, Austria, Canada, Croatia, France, Greece, Germany, Italy, Japan, Mexico, Portugal, Spain and Switzerland.
Product codes and lots
Lot Numbers ED2098, EA2111, FB2134, ED2147