Z-0086-2012

O'Neil Sterile Field Intermittent Urinary Catheter Kit labeled in mart:"***ORIGINAL CLOSED SYSTEM BY DR...

Go Medical Industries Pty., Ltd. / initiated 2011-09-03 / Terminated / root cause: manufacturing process

Go Medical Industries Pty., Ltd. began a recall of O'Neil Sterile Field Intermittent Urinary Catheter Kit labeled in mart:"***ORIGINAL CLOSED SYSTEM BY DR O'NEIL O'NEIL STERILE FIELD INTERMITTENT URINARY... on . The reason FDA records is "O'Neil Sterile Field Intermittent Urinary Catheter Kits With Povidone Iodine Swabsticks are being recalled due to a recall on the Triad Povidone Iodine Swabsticks contained within the kit. H&P Industries are recalling...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0086-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Go Medical Industries Pty., Ltd.
Product
O'Neil Sterile Field Intermittent Urinary Catheter Kit labeled in mart:"***ORIGINAL CLOSED SYSTEM BY DR O'NEIL O'NEIL STERILE FIELD INTERMITTENT URINARY CATHETER KIT WITH IODINE SWABSTICKS (3)*** REF: ONSM 8KPES SIZE: 8Fr Product No: AAM 8008KPES***SINGLE USE ONLY***Original design by: Dr. George O'Neil 200 Churchill Ave, Subiaco, WA 6008, Australia Tel.: +61 (0)8 9388 1700 Fax: +61 (0)8 9382 4952 European Representative: Advena Ltd Hereford, HR4 9DQ, UK DISTRIBUTED IN THE USA BY: American Australian Medical, Inc. 1-888-880-8871***" Case of Catheter Kits labeled in part:"***The O'Neil Sterile Field Urinary Catheter Kit REF: ONSM 8KPES Product No: AAM 8008KPES***CAUTION: Federal law restrict sale of this device to or on the order of a physician. STERILE Single use only***Designed by: GO MEDICAL INDUSTRIES PTY LTD 200 Churchill Avenue, Subiaco, WA 6008, Australia Tel.: +61 (0)8 9388 1700 Fax: +61 (0)8 9382 4952 email: [email protected] Website: www.gomedical.com.au***Distributed in the U.S.A. by: AMERICAN AUSTRALIAN MEDICAL INC. BUFFALO, NY 14209 USA TEL: 1-888-880-8871***Country of Origin: ROMANIA***QUANTITY 50***" This kit includes a Urinary Catheter and accessories including gloves, tissue, underpads, and Triad Povidone Iodine Swabsticks (3). These accessories are included in the kit as optional and are not required for urinary catheterization.
Product code
FCM Tray, Catheterization, Sterile Urethral, With Or Without Catheter (Kit)
Reason for recall
O'Neil Sterile Field Intermittent Urinary Catheter Kits With Povidone Iodine Swabsticks are being recalled due to a recall on the Triad Povidone Iodine Swabsticks contained within the kit. H&P Industries are recalling the Triad Povidone Iodine Swabsticks because it was determined that the swabsticks were manufactured without having a system for microbial testing in place at the time of release.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
The firm, Go Medical Industries Pty Ltd, sent an e-mail on September 3, 2011 and follow-up with an "URGENT: MEDICAL DEVICE RECALL" letter dated September 07, 2011 to its customers. The letter described the product, problem and action to be taken. The customers were instructed to quarantine and not sell the product in the market place; immediately perform a sub-recall and notify their customers and complete and return the RETURN RESPONSE FORM via fax at 716-883-8099 or mail to: American Australian Medical Inc, 1083 Delaware Avenue, Buffalo, NY 14209. For further details you can contact QA Manager at Go Medical Industries at +61 89 381 3226 or [email protected].
Quantity in commerce
6,300
Distribution
Nationwide distribution.
Product codes and lots
Lot #'s: 9430 9431 9432