Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0054-2012

Sarns Intracardiac Sucker, Bulk, Non-Sterile

Terumo Cardiovascular Systems Corporation / initiated 2011-09-13 / Terminated / root cause: manufacturing process

Terumo Cardiovascular Systems Corporation began a recall of SARNS INTRACARDIAC SUCKER, BULK, NON-STERILE on . The reason FDA records is "Terumo Cardiovascular Systems (Terumo CVS) recently discovered three instances of metal fragments inside the stainless steel tube of the Sarns" Adult Rigid Intracardiac Sucker". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0054-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
Sarns Intracardiac Sucker, Bulk, Non-Sterile
Product code
DTS Sucker, Cardiotomy Return, Cardiopulmonary Bypass
Reason for recall
Terumo Cardiovascular Systems (Terumo CVS) recently discovered three instances of metal fragments inside the stainless steel tube of the Sarns" Adult Rigid Intracardiac Sucker.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
On 9/13/2011 Terumo sent a letter entitled "URGENT MEDICAL DEVICE REMOVAL" to its customers that included the name and description of the device, lot numbers, the potential hazard, correction, the reason for the recall and instructions. Customers are to review the notice, ensure that all users are aware of the issue, and a recall response form to be completed and returned. Customers are to call Terumo CVS Customer Service to obtain a Return Goods Authorization number. The letter also included MedWatch reporting information and stated that Customers with questions should call 1-800-521-2818.
Quantity in commerce
5700
Distribution
Worldwide Distribution -- USA, Belgium, Australia, Mexico, and Japan.
Product codes and lots
Catalog number 11009010