Z-0932-2012

Solanas Posterior Stabilization System Part Number: 63920. Solanas Posterior Stabilization System facilitates...

Alphatec Spine, Inc. / initiated 2010-11-01 / Terminated / root cause: design

Alphatec Spine, Inc. began a recall of Solanas Posterior Stabilization System Part Number: 63920. Solanas Posterior Stabilization System facilitates the surgical correction of spinal deformities by... on . The reason FDA records is "These lots were not manufactured correctly in that the instrument does not have the durability required to repeatedly perform over its full service life and, therefore, may experience breakage in the tips". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0932-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Alphatec Spine, Inc.
Product
Solanas Posterior Stabilization System Part Number: 63920. Solanas Posterior Stabilization System facilitates the surgical correction of spinal deformities by providing temporary internal fixation and stabilization during bone graft healing and/or fusion mass development. The Solanas Posterior Stabilization System is a spinal fixation system intended to improve stability of the cervical and thoracolumbar area of the spine (C1-T3).
Product code
KWP Appliance, Fixation, Spinal Interlaminal
Reason for recall
These lots were not manufactured correctly in that the instrument does not have the durability required to repeatedly perform over its full service life and, therefore, may experience breakage in the tips.
Root cause tag
design
FDA root cause
Device Design
Action
Alphatec Spine contacted all affected customers via telephone. The communication included discussion of the product, problem, and actions to be taken by the customers. Customers were instructed to remove the affected product from inventory and to return the product to the firm. Arrangements were made for replacement product to be sent to the customers. Contact the firm at 1-800-922-1356 for questions regarding this recall.
Quantity in commerce
27 units
Distribution
Nationwide Distribution-including the states of CA, CT, FL, GA, NC, NE, NJ, TX, and UT.
Product codes and lots
Lot Numbers: 608227B, 609867B, 609870B, 609882B, 609897B, 609971B