CareFusion EnVe Ventilator, Model 19250-001. Manufactured by: CareFusion, 22745 Savi Ranch, Yorba Linda, CA...
CareFusion 203, Inc. began a recall of CareFusion EnVe Ventilator, Model 19250-001. Manufactured by: CareFusion, 22745 Savi Ranch, Yorba Linda, CA 92887 USA. EnVe is a portable critical care... on . The reason FDA records is "CareFusion has identified potential risks associated with the EnVe ventilator manufactured between December 2010 and May 2011. There are 3 issues that can result in ventilation delivery to the patient being interrupted...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0074-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Pulmonetic Systems, Inc.
- Product
- CareFusion EnVe Ventilator, Model 19250-001. Manufactured by: CareFusion, 22745 Savi Ranch, Yorba Linda, CA 92887 USA. EnVe is a portable critical care ventilator. It is suitable for a wide range of life support and critical care applications. The ventilator normally operates from external DC power. When the ventilator is portable, a Removable Battery Pack powers the unit. Devices are designed for use on patients who require respiratory support or mechanical ventilation and weigh a minimum of 5 kg (11 lbs).
- Product code
- CBK Ventilator, Continuous, Facility Use
- Reason for recall
- CareFusion has identified potential risks associated with the EnVe ventilator manufactured between December 2010 and May 2011. There are 3 issues that can result in ventilation delivery to the patient being interrupted. CareFusion will be contacting facilities to coordinate implementation of the necessary corrective actions.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- The firm, CareFusion, sent an "URGENT PRODUCT RECALL" letter dated September 12, 2011 to its customers.The letter identified affected product, described the 3 issues providing the description of the event, potential risk, and provided required action for Users. The letter also stated that CareFusion would be contacting the facility to coordinate the corrective action. If you have any questions and/or need support, please contact CareFusion Technical Support at 1-800-554-8933 from Mon. - Fri. 8:00am -5:00pm Pacific Time.
- Quantity in commerce
- 128 units
- Distribution
- Nationwide distribution: USA including states of: CA, CO, FL, IL, MI, MN, MT, ND, NJ, NM, NY, OK, PA, TX, UT, WI, and WV.
- Product codes and lots
- serial numbers: 45317, 45318, 45319, 45321, 45323, 45324, 45326, 45327, 45328, 45330, 45331, 45332, 45333, 45334, 45336, 45338, 45339, 45341, 45735, 45736, 45737, 45738, 45740, 45741, 45742, 45744, 45745, 45746, 45747, 45748, 45749, 45750, 45751, 45752, 45754, 45902, 45905, 45906, 45907, 45908, 45909, 45910, 46038, 46039, 46040, 46041, 46042, 46043, 46044, 46046, 46050, 46051, 46632, 46634, 46635, 46636, 46637, 46638, 46640, 46641, 46642, 46643, 46644, 46645, 47069, 47070, 47071, 47073, 47074, 47075, 47076, 47077, 47078, 47079, 47081, 47082, 47083, 47210, 47211, 47213, 47214, 47215, 47216, 47217, 47219, 47221, 47222, 47223, 47224, 47225, 47226, 47229, 47230, 47231, 47233, 47234, 47236, 47237, 47238, 47239, 47538, 47539, 47540, 47541, 47542, 47544, 47545, 47546, 47547, 47548, 47549, 47550, 47551, 47552, 47553, 47554, 47555, 47556, 47557, 47558, 47559, 47560, 47561, 47562, 47563, 47564, 47566, 47567.